Specifications

 

 

 

 

 

 

 

Intake Purge

Air purged automatically before and after each sample; duration automatically compensates

 

 

for varying intake line lengths.

 

 

 

 

 

 

 

 

 

 

 

Pump/Controller

High impact injection molded ABS; submersible, watertight, dust tight, corrosion & ice

 

 

Housing

resistant; NEMA 4X,6.

 

 

 

 

 

 

 

Internal Clock

Indicates real time and date; 0.007% time base accuracy.

 

 

 

 

 

 

 

Manual Sample

Initiates a sample collection independent of program in progress.

 

 

 

 

 

 

 

Intake Rinse

Intake line automatically rinsed with source liquid prior to each sample, from 0 to 3 rinses.

 

 

 

 

 

 

 

Intake Retries

Sample collection cycle automatically repeated from 0 to 3 times if sample is not obtained on

 

 

initial attempt.

 

 

 

 

 

 

 

 

 

 

 

Intake Tubing

¼ in. and 3/8 in. ID vinyl or 3/8 in. ID Teflon* lined polyethylene with protective outer cover.

 

 

Intake Strainers

Choice of Teflon® and 316 stainless construction, and all 316 stainless steel in standard size

 

 

and low profile for shallow depth applications.

 

 

 

 

 

 

 

 

 

 

Factory Installed Options

 

 

 

 

 

 

pH/Temperature/ORP Meter

 

 

 

 

 

 

 

 

Control/Logging

 

Field selectable to log pH/temperature or ORP independent of sample operation or to control

 

 

 

sample collection in response to volume exceeding low/high setpoints.

 

 

 

 

 

 

 

 

 

 

 

pH/Temperature Sensor

 

Temperature compensated; impact resistant ABS plastic body

 

 

 

Combination electrode with porous Teflon junction

 

 

 

 

 

 

 

 

 

 

 

Measurement Range

 

0 to 14 pH, -10 to 105 °C

 

 

 

 

 

 

 

Operating Temperature

 

-18 to 80 °C (0 to 176 °F)

 

 

 

 

 

 

 

Dimensions

 

1.9 cm dia. 15.2 cm long (0.75 in. 6 in.) with 1.9 cm (0.75 in.) mpt cable end

 

 

 

 

 

 

Dissolved Oxygen Meter

 

 

 

 

 

 

 

 

Control/Logging

 

Field selectable to log dissolved oxygen independent of sampler operation or to control

 

 

 

sample collection in response to volume exceeding low/high setpoints.

 

 

 

 

 

 

 

 

 

 

 

Measurement Method

 

Galvanic

 

 

 

 

 

 

 

Sensor

 

Temperature compensated; impact resistant polypropylene body

 

 

 

 

 

 

 

Measurement Range

 

0 to 20 mg/L

 

 

 

 

 

 

 

Resolution

 

0.01 mg/L

 

 

 

 

 

 

 

Accuracy

 

±3% of reading or 0.1 mg/L

 

 

 

 

 

 

 

Operating Temperature

 

0 to 50 °C (32 to 122 °F)

 

 

 

 

 

 

 

Dimensions

 

1.7 mm diameter 15.7 cm long (0.65 in. 6.25 in.) with 1.9 cm (0.75 in.)

 

 

 

 

 

 

Conductivity Meter

 

 

 

 

 

 

 

 

Control/Logging

 

Field selectable to log conductivity independent of sampler connection or to control sample

 

 

 

collection in response to volume exceeding low/high setpoints.

 

 

 

 

 

 

 

 

 

 

 

Sensor

 

Temperature compensated; impact resistant polypropylene body

 

 

 

 

 

 

 

Measurement Range

 

0 to 20 mS/cm

 

 

 

 

 

 

 

Resolution

 

0.01 mS/cm or 1 mS/cm

 

 

 

 

 

 

 

Accuracy

 

±2% of reading or 0.01 mS

 

 

 

 

 

 

 

Operating Temperature

 

0 to 50 °C (32 to 122 °F)

 

 

 

 

 

 

 

Dimensions

 

1.7 cm diameter 15.2 cm long (0.67 in. x 6 in.) with 1.9 cm (0.75 in.) mpt cable end

 

 

 

 

 

 

 

 

 

 

 

 

 

 