Section 5

5.5.4Conductivity Probe Calibration

1.From the Main Menu, select OPTIONS > ADVANCED OPTIONS >

CALIBRATION > CONDUCTIVITY.

2.Clean and dry the probe.

3.Place the sensor and thermometer in the calibration solution

(Cat. No. 3230). The temperature sensor is located in the middle of the sensor body allowing the probe to be completely submerged in

the solution.

4.Allow the sensor to stabilize in the solution about 10 minutes to ensure that the probe and the solution are the same temperature.

5.Enter the temperature correction factor or enter zero for no correction factor.

Note: The temperature correction factor is used to compensate for the effects of temperature on the conductivity readings at the point of installation. The conductivity of a solution is temperature sensitive. Therefore the actual conductivity of the solution will change with the temperature. Each site may have a different correction factor depending on the major constituent of the flow stream. This is not used for calibration and has no effect on the calibration of the sensor. Below are some examples of compensation factors of various liquids.

0.96%/°C 5% Sulfuric Acid

1.88%/°C Dilute Ammonia

1.91%/°C ‘Typical’ Wastewater

1.97%/°C Potassium Chloride

2.12%/°C Salt (Sodium Chloride)

2.84%/°C 98% Sulfuric Acid

4.55%/°C Ultra-pure Water

6.With the sensor still in the calibration solution, press any key. Wait for the sensor to stabilize. Calculate the actual conductivity of the calibration solution. If using the KCl solution provided by the manufacturer, make your selection from the Table 12 on page 78. If using a solution other than 1.0 mS @ 25 °C KCl available from manufacturer, calculate the conductivity of the solution using temperature correction factors. See the following example.

Example: The KCl calibration solution is 1.0 mS at 25°C. If the actual temperature of the KCl at the time of calibration is 18.4 °C, then the solution has a conductivity value of 0.870 mS.

a.Find the difference between the labeled temperature and the actual temperature of the calibration solution at the time of calibration.

25 °C – 18.4 °C = 6.6 °C

b.Multiply the difference (6.6) by the correction factor per °C (1.97% or 0.0197).

6.6°C x 0.0197/°C = 0.13002

 

Page 77

8990options.fm

Conductivity Probe

Page 79
Image 79
Hach 900 MAX manual Conductivity Probe Calibration

900 MAX specifications

The Hach 900 MAX is an advanced analytical instrument designed to perform water quality analysis with precision and efficiency. This multiparameter water quality analyzer is tailored for a wide range of applications, including environmental monitoring, wastewater treatment, and industrial processes. Its unique design and technology make it a valuable tool for professionals in the field.

One of the standout features of the Hach 900 MAX is its ability to measure multiple parameters simultaneously. Equipped with advanced sensors, the device can analyze key water quality indicators such as pH, turbidity, dissolved oxygen, conductivity, and various chemical parameters. This comprehensive measurement capability allows users to gain a complete understanding of water quality in a single run, saving both time and resources.

The Hach 900 MAX utilizes cutting-edge technologies to provide accurate and reliable results. Its integrated optical sensors employ advanced photometry, which enhances the detection of chemicals in water samples with high sensitivity. Additionally, the device includes advanced calibration management tools that ensure consistent accuracy over time. Users can easily perform routine calibrations with minimal effort, facilitating reliable data collection.

Another notable characteristic of the Hach 900 MAX is its user-friendly interface. The instrument features a large, touchscreen display that allows for intuitive navigation through various settings and parameters. Users can quickly access real-time data, generate reports, and analyze trends using the built-in software. The ability to customize dashboards and create visual representations of data enhances user experience and simplifies the decision-making process.

In addition to its analytical capabilities, the Hach 900 MAX emphasizes connectivity and data management. The instrument can be connected to laboratory information management systems (LIMS) and other data networks, enabling smooth data transfer and integration into existing workflows. This connectivity enhances collaboration among team members and facilitates compliance with regulatory requirements.

Durability is another hallmark of the Hach 900 MAX. The device is designed for use in demanding environments, with robust construction and protection against dust and moisture. This reliability ensures that the analyzer remains operational even in challenging conditions, making it suitable for field applications.

Overall, the Hach 900 MAX stands out as a powerful and versatile tool in the realm of water quality analysis. With its multiparameter measurement capabilities, advanced sensor technologies, user-friendly interface, and reliable performance, it enables professionals to obtain accurate insights into water quality efficiently. As environmental concerns continue to grow, instruments like the Hach 900 MAX play a crucial role in ensuring the safety and sustainability of water resources.