Hach 900 MAX manual Section, Sampler Pacing Flow Units

Models: 900 MAX

1 160
Download 160 pages 21.13 Kb
Page 44
Image 44
Table 2 Sampler Pacing Flow Units

Section 3

c.Select either Integral or External flow meter and press ACCEPT.

d.Enter the flow volume between samples using the numeric keypad and select a unit of measure using the CHANGE UNITS key, then press ACCEPT. Refer to Table 2 for flow unit choices.

11:00 AM 21 - APR - 01

 

SAMPLE COLLECTION

ACCEPT

 

CHANGE

TAKE SAMPLE EVERY:

UNITS

 

 

1500 gal

 

CANCEL

 

CLEAR

 

ENTRY

 

 

 

 

 

 

ENTER: 1.00—99999999

 

 

 

 

 

Table 2 Sampler Pacing Flow Units

 

 

 

Abbreviation

 

Volume

 

 

 

gal

 

gallons

 

 

 

ltr

 

liters

 

 

 

m3

 

cubic meters

af

 

acre-feet

 

 

 

cf

 

cubic feet

 

 

 

e.Enable or Disable Timed Over-Ride using the CHANGE CHOICE key. Press ACCEPT to continue, then enter a time period using the numeric keypad.

Note: Select Timed Over-Ride if the flow rate drops to an unusually low value during flow-proportional sampling and if the sample is collected once an hour for example, even if the flow interval has not elapsed.

11:00 AM 21 - APR - 01

SAMPLE COLLECTION

 

 

ACCEPT

TIMED OVER-RIDE: 00:00 (hrs:min)

CLEAR

CANCEL

ENTRY

ENTER: 000:01 — 999:00 (hrs:min)

f.Select Take First Sample Immediately or After First Interval?

Note: When the program is started, the first sample is taken immediately upon pressing the start button or after the first interval has elapsed.

11:00 AM 21 - APR - 01

 

SAMPLE COLLECTION

 

ACCEPT

 

CHANGE

 

CHOICE

 

 

 

TAKE FIRST SAMPLE:

 

_______________

CANCEL

 

 

 

 

 

CHOICES: IMMEDIATE, AFTER 1st SAMPLE

g.Press ACCEPT to continue to Sample Distribution.

Page 42

 

Basic Programming Setup

8990startup.fm