Hach 900 MAX manual Special Output, Section, Sampling Triggers and Settings

Models: 900 MAX

1 160
Download 160 pages 21.13 Kb
Page 54
Image 54
Table 3 Sampling Triggers and Settings

Section 3

13-I.Enter a delay when input is active. This delay will keep the program from starting until the end of the delay period. Using the numeric keypad, enter the delay in minutes and/or hours. Press ACCEPT.

11:00 AM 21 - APR - 01

SETPOINT SAMPLING

ACCEPT

DELAY WHEN INPUT

 

 

BECOMES ACTIVE:

 

CANCEL

(hrs:min)

CLEAR

 

ENTRY

 

 

 

 

 

 

(USE NUMERIC KEYPAD)

 

 

 

 

 

Table 3 Sampling Triggers and Settings

 

 

 

Channel

Sampling Trigger

Settings

 

 

 

1

Level

High and/or Low Condition, Deadband

 

 

 

2

Flow

High and/or Low Condition, Deadband

 

 

 

3

Flow Rate of Change

High Condition within Time Interval

 

 

 

4

pH or ORP

High and/or Low Condition, Deadband

 

 

 

5

Process Temperature

High and/or Low Condition, Deadband

 

 

 

6

Rainfall

High Condition within Timed Interval

 

 

 

7

Analog Input Channel 1

High and/or Low Condition, Deadband

 

 

 

8

Analog Input Channel 2

High and/or Low Condition, Deadband

 

 

 

9

Analog Input Channel 3

High and/or Low Condition, Deadband

 

 

 

10

Analog Input Channel 4 or DO

High and/or Low Condition, Deadband

 

 

 

11

Analog Input Channel 5 or DO Temperature

High and/or Low Condition, Deadband

 

 

 

12

Analog Input Channel 6 or Conductivity

High and/or Low Condition, Deadband

 

 

 

13

Analog Input Channel 7 or Conductivity Temperature

High and/or Low Condition, Deadband

 

 

 

14

External Control

Configured in External Equipment

 

 

 

Step 14 - Special Output

The Special Output is a +12 V dc signal that appears on Pin E of the Auxiliary

Receptacle (refer to section 2.10 on page 29).

14-A.Highlight Special Output using the UP and DOWN keys on the Advanced Sampling Menu. Press SELECT to continue.

11:00 AM 21 - APR - 01

ADVANCED SAMPLING

SELECT

RETURN

PGM CMPLT OUTPUT SETPOINT SAMPLING

SPECIAL OUTPUTS

START/STOP TIMES STORM WATER

14-B.Enable or Disable Special Outputs using the CHANGE CHOICE key. Press SELECT to continue.

14-C.If enabled, select After Each Sample, Only When Pumping, or From Rinse to Purge.

Page 52

 

Advanced Sampling

8990startup.fm

Page 54
Image 54
Hach 900 MAX manual Special Output, Section, Sampling Triggers and Settings