Hach 900 MAX manual Appendix B, Sensor Enabled, Calibration Volumes

Models: 900 MAX

1 160
Download 160 pages 21.13 Kb
Page 119
Image 119
Appendix B

Appendix B

Sensor Enabled

When the sensor is enabled, and volume calibration is selected, a single manual calibration is taken. It does not matter if multiple volumes are programmed (such as when using stormwater or upset sampling). When the sensor is enabled, a single “Manual Calibration” choice provides sufficient information to adjust the sensor to all programmed volumes (see Table 24).

 

Table 24 Calibration Volumes

 

 

 

Liquid Sampler

 

Available Calibration Volumes

 

 

 

Enabled

 

Manual Calibration

 

 

 

Disabled

 

Main Sample Volume

 

 

 

Disabled/Stormwater

 

Main Sample Volume

 

 

 

First Flush Volume

 

 

 

 

 

Disabled/Upset Sampling

 

Main Sample Volume

 

 

 

Upset Sample Volume

 

 

 

 

 

 

 

Main Sample Volume

 

 

 

Disabled/Storm & Upset

 

Upset Sample Volume

 

 

 

 

 

First flush Volume

 

 

 

Disabled/If Timed Rinses Set

 

Timed Rinse (in addition to the above)

 

 

 

Note: Before calibrating, make certain the desired sample volume in the Setup menu is selected.

Calibration Procedure—Sensor Disabled:

1.From the Main Menu, press OPTIONS.

2.Press VOLUME CALIBRATION.

3.Select the desired volume to calibrate if more than one is listed (see Table 24).

4.Slip the pump tube off the sample fitting (near where it exits the pump housing) and direct the end of the pump tube into a graduated cylinder.

5.Press START PUMPING. The pump will purge the line, and start pumping a sample.

6.Press any key when the desired sample volume has been reached in the graduated cylinder. The pump will stop to allow the line to completely drain and then reverse to post-purge the line.

7.If the pumped sample volume is correct, press DONE. If not, repeat the calibration cycle by pressing TRY AGAIN.

8.When finished, re-install the pump tube onto the sample fitting.

Note: To check the sample size after calibration use the Manual Mode function key to take a grab sample. Do not go back into calibration to check the volume since the volume compensation is reset to zero at the start of a calibration.

If one or more Timed Rinses are selected (sensor disabled), then they must be calibrated as well. The Timed Rinse will rinse the intake line up to 3 times prior to taking a sample.

 

Page 117

8990apdx_pfeatures.fm

Programming Features

Page 119
Image 119
Hach 900 MAX manual Appendix B, Sensor Enabled, Calibration Volumes, Calibration Procedure—SensorDisabled