Starter Kit Instruction Manual

Use

34

6.5Interlock utilisation

The temperature controller is designed with an built- in interlock (1). This function is activated when a fault occurs on the TCU151 controller (over-temperature, over-current, etc.).

The interlock is by default a NC (Normally Close) relay contact.

CAUTION ! It is also possible to set the Inter- lock as a NO (Normally Open) contact. For more details about the settings, see the procedure on page 42.

6.5.1Before beginning

:

 

1

M

 

o

 

h

0

m

 

1

 

r

 

l

o

 

c

 

ki

 

n

 

g

1

 

Fig.24: Interlock BNC connector

In order to recover the temperature controller from an interlock event, please pay attention about the following points:

Locate the fault which caused the interlock and fix it.

Wait until the temperature driven by the TCU151 is lower than the maximum.

6.5.2Procedure Interlock utilisation

To reset the temperature controller from an inter- lock event, proceed as follows:

[1]Press the RESET button (3) on the TCU151 front panel

,→ The interlock is reset and the red LED ALARM (2) should be off.

Al

 

ar

 

m

 

R

 

e

 

s

 

et

E

 

 

xt

 

er

 

n

 

al

2 Internal

3

Fig.25: Reseting the interlock

Page 35
Image 35
Quantum Instruments Cascade Laser Starter Kit manual Procedure Interlock utilisation