Starter Kit Instruction Manual

Description

19

4.4.4Measurement connection

These connections give access to the voltage on the laser.

The ”L” connection is connected AC coupled to the cathode of the laser through a divider by ten. The end of cable must be 50Ω terminated for accu- rate measurement. The ”B” is connected the same to the anode of the laser (Base Receptacle).

By using the math trace of an oscilloscope show- ing the ”L” channel minus the ”B”, one can get the voltage on the laser.

This measurement is more accurate than the in- formation obtained from the LDD100 measurement unit.

For more details, refer to LDD100 Diagram Block on page ??.

Specifications of the measurement connector

Vacuum case plug reference: . . . . . . . LEMO #

HGP.00.250.CTLPV

Cable plug reference: . . . . . . . . . . . . . . LEMO #

FFS.00.250.CTCE31

Cable : . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . RG174