Starter Kit Instruction Manual | Description | 19 |
4.4.4Measurement connection
These connections give access to the voltage on the laser.
The ”L” connection is connected AC coupled to the cathode of the laser through a divider by ten. The end of cable must be 50Ω terminated for accu- rate measurement. The ”B” is connected the same to the anode of the laser (Base Receptacle).
By using the math trace of an oscilloscope show- ing the ”L” channel minus the ”B”, one can get the voltage on the laser.
This measurement is more accurate than the in- formation obtained from the LDD100 measurement unit.
For more details, refer to LDD100 Diagram Block on page ??.
Specifications of the measurement connector
Vacuum case plug reference: . . . . . . . LEMO #
HGP.00.250.CTLPV
Cable plug reference: . . . . . . . . . . . . . . LEMO #
FFS.00.250.CTCE31
Cable : . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . RG174