Starter Kit Instruction Manual

Maintenance

40

8.2.2 QCL ”UP” and ”DN” position

 

 

exchange

 

 

 

Before beginning

 

 

 

[1] Switch the QCL Starter Kit OFF.

 

1

 

 

 

 

Material needed

 

 

 

• 1 set of Allen keys.

 

 

 

• 1 ohmmeter.

 

 

 

• 1 FP-QCL laser contact plate.

 

 

 

• 1 pair of tweezers.

 

2

 

 

 

 

Procedure

 

 

 

To exchange the beam of a Quantum Cascade Laser,

 

 

proceed as follows:

 

3

 

 

 

 

[1] Remove the top cover (2) of the Laboratory

4

 

Laser Housing (LLH100) by unscrewing the

 

screws (1).

 

 

 

[2] Remove the PET U-shaped holder (4) by

 

 

unscrewing the fixing screws (3).

 

 

 

Caution ! Pay attention NOT to drop the

 

 

laser!

 

 

 

[3] Remove the gilded copper contact (6) by

 

 

unscrewing the back screw (5).

 

 

 

[4] Install the opposite gilded copper contact Fig.28: Removing the laser recptacle

(6) by screwing back the screw (5).

5

6

Fig.29: Exchanging the laser

Page 41
Image 41
Quantum Instruments Cascade Laser Starter Kit manual QCL UP and DN position Exchange, Exchanging the laser