Starter Kit Instruction Manual  | Maintenance  | 39  | |
[4] Install the new laser (4) into the holder  | 
  | 
  | |
using small tweezers ( 0.5cm aperture).  | 
  | 
  | 
  | 
[5] Hold the laser (4) on the gilded ceramic  | 1  | 
  | |
  | 
  | ||
pads with the tweezers pins along the axis of  | 
  | 
  | |
the largest length of the copper submount.  | 
  | 
  | 
  | 
[6] Place it in the receptacle parallel to the  | 
  | 
  | |
window.  | 
  | 
  | 
  | 
! Warning ! Avoid hitting the laser chip  | 2  | 
  | |
facet on the case, it is extremely fragile. | 
  | ||
[7] Place the screws (3) in the PET holder (5).  | 3  | 5  | |
[8] Place the PET holder (5) on the two guides  | 
  | ||
  | 
  | ||
(6) on the back of the laser receptacle being  | 
  | 
  | |
careful of maintaining the PET holder higher  | 
  | 
  | |
on the laser side than on the back side.  | 
  | 
  | 6  | 
  | 
  | 
  | |
[9] Once the PET holder (5) is engaged in the  | 4  | 
  | |
guides, push it down until contact with the  | 
  | 
  | |
gilded ceramic pads is established.  | 
  | 
  | 
  | 
[10] Screw it to the receptacle.  | 
  | 
  | 
  | 
[11] Verify that the contact is established by  | 
  | 
  | |
measuring with an ohm meter on the LLH100  | 
  | 
  | |
low impedance plug.  | 
  | 
  | 
  | 
Note: For reproducible measurement be care-  | 
  | 
  | |
ful of measuring with the positive probe on the  | Fig.27: Replacing the laser | 
  | |
+ LLH100 connection. | 
  | 
  | 
  | 
Note: If the contact is bad, adjust the plastic screws at the pressing end of the PET holder
(5).
Note: If the contact is bad, clean the bottom of the submount with grinding paper, be care- ful of not damaging the bonding wires and the laser facet.
[12]Place the dry desiccant bag close to the low impedance line plug. Only applicable if desiccation is needed and the lasers has been shipped with a desiccant bag.
Note: This operation is preferably performed in a gloves box under dry nitrogen atmosphere if the ambient atmosphere is humid (this is not necessary under normal conditions 
[13] Close the cover (2) with the screws (1).