Starter Kit Instruction Manual Description 25
4.8 QCL temp erature controller(TCU151)

4.8.1 Description

The TCU151 is used to control the laser’s temper-
ature inside the Laboratory Laser Housing.
It uses a PT100 sensor to measure the tempera-
ture of the cold plate and maintains a pre-set tem-
perature either from a front panel knob or from a
user supplied voltage.

4.8.2 Specifications

Voltage:. . . . . . . . . . . . . . . . . . . . . . . 120-240V AC
Frequency:. . . . . . .. . . . . . . . . . . . . . . . . . 50-60 Hz
Fuse:
.. . . . . . . . . . . . . . . . . . . 1.6 A T (220-240V)
.. . . . . . . . . . . . . . . . . . 3.2 A T (110 -12 0V)
Peltier current maximum: . . . . . . . . . . . . . .. . 5A
Temperature control range: . -65 C to 65 C
Temperature measurement: . . . PT100 4 wires
Temperature monitor: . .. . 10mV/C, Zout =
200Ω
In addition, allow for 15 min. warm-up of
TCU151 to get stable readings
External driving temp erature: . -6.5V to 6.5V
(-65 C to 65 C), Zin =1MΩ

4.8.3 Peltier and PT100 connections

The rear connector pinout is listed according to the
following items:
(1) Pin 1: +Peltier element
(2) Pin 2: - Peltier element
(3) Pin 3: +I PT100
(4) Pin 4: +Sense PT100
(5) Pin 5: -Sense PT100
(6) Pin 6: -I PT100
(7) Pin GND: Ground
TCU151
T
emperatureController
AlarmResetExternal
Internal
RealI
RealC
SettingC
Setting+I
Setting−I
+
Fig.11: TCU151 temperature controller
1
2
1
7
3
4
5
6
Fig.12: TCU151 Peltier Connector