Starter Kit Instruction Manual

Maintenance

43

8.4Calibaration procedures

8.4.1TPG128 calibration

Generalities

The present section describes procedure to be car- ried out for particular calibration of the TPG128.

Before beginning

[1]Switch the instrument OFF.

[2]Remove the cover (1).

[3]Switch the instrument ON.

!WARNING ! Keep in mind that dur- ing the calibration, the instrument is alive (powered on)

Needed material

1 set of screwdriver.

1 oscilloscope.

1 voltmeter.

1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O

FF

 

T

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

P

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

G

12

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

w

N

 

 

 

 

 

8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

T

 

 

 

 

 

 

 

 

 

 

 

 

 

 

er

 

 

 

 

 

 

 

 

 

TL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

−−

 

 

 

 

 

 

Pu

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

l

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.

 

se

 

 

 

 

 

 

 

 

 

 

+

 

 

 

−−

 

 

 

 

 

 

 

Ge

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2 t

 

 

 

 

 

 

 

max

 

12

 

 

−−

 

 

 

0.5 o

2.

ne

 

 

 

 

 

 

0

 

V

 

 

 

 

 

 

5

 

to

 

2

 

ra

 

 

 

 

 

60mA

 

V

 

 

 

 

 

Perio

 

to

1010.5ms

 

tor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

d

 

 

 

5

 

ms

 

0 t

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ms

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

−−

 

 

 

 

 

 

o

20

 

 

 

 

 

 

 

 

 

 

 

 

 

50

oh

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Gat

 

 

 

 

 

−−

 

 

 

 

 

 

0n

 

 

 

 

 

 

 

 

 

 

 

 

ms

 

 

−−

 

 

 

 

 

s

 

 

 

 

 

 

 

 

eI

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N

 

Tri

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

50 oh

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

g O

 

 

 

 

 

 

ms

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

U

 

 

 

 

 

 

r

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

T

 

 

 

 

 

ati

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O

 

50 oh

 

on

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ms

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ut

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ut

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

 

 

 

Procedure calibration

The generator should be calibrated as followed:

[1]Verify the +5 V supply.

[2]During the first test, set all potentiometers and adjustable capacitors to their medium val- ues.

[3]Adjust to the potentiometer 50 ns adj

(3)in order to obtain a 500 ns long pulse on TP2 (2).

[4]Connect an oscilloscope allowing the mea- surement of the signal’s period on the output Trig out.

Fig.33: Removing the top cover

2

3

Fig.34: TPG128 main board

Page 44
Image 44
Quantum Instruments Cascade Laser Starter Kit Calibaration procedures, 1 TPG128 calibration, Generalities, Needed material