Starter Kit Instruction Manual Maintenance 42
8.3 T CU151 temperature con-troller interlock setting

8.3.1 Generalities

The interlock is located on the main board.
Note: By default, the interlock is set to NC (Nor-
mally Close) contact.

8.3.2 Before beggining interlock set-

ting

[1] Switch the instrument OFF.
[2] Unplug the power cord.
[3] Remove the cover (1) to acces the main
board (2).

8.3.3 Needed material

1 set of screwdriver.
1 tweezer.

8.3.4 Procedure interlock setting

To set the type of interlock, proceed as follows:
[1] Locate the J9 Interlock connector.
[2] Place the associated jumper according to
the type of desired interlock.
Note: The common pin is on the center (4)
and by placing the jumper on a side defines
the interlock type as:
for a NC contact: jumper on pins (4) and
(5)
for a NO contact: jumper on pins (3)
and (4)
2
1
T
emperatureController
AlarmResetExternal
Internal
R
ealI
R
ealC
SettingC
S
etting+I
S
etting
I
+
TCU151
Fig.31: Opening the TCU151 cover
345
Fig.32: J9 Interlock connector