4.4.3Thermo-Electric Cooler (TEC) and PT100 connections

The pin 1 is recognised by a circle around it and a thin noon ended circle starts from pin one and turns around the pins until pin 6.

(1)

Pin 1: +Peltier element

(2)

Pin 2: - Peltier element

(3)

Pin 3: I+ PT100

6

1

 

5

(4)

Pin 4: Sens + PT100

(5)

Pin 5: Sens - PT100

(6)

Pin 6: I- PT100

Specifications of the Peltier connector

4

2

3

Vacuum case plug reference: . . . . . . . LEMO # HGG.1b.306.CLLPV.

Cable plug reference: . . . . . . . . . . . . . LEMO # FGG.1b.306.CLAD72Z.

Plastic protection for the cable: . . .LEMO # GMA.1b.065.DG.

Cable:

Diameter : . . . . . . . . . . . . . . . . . . . . . 5.8 mm

Section : 6 x 0.25 mm (2 x 0.25 for each Peltier connections)

or

Diameter : . . . . . . . . . . . . . . . . . . . . . 7.2 mm

Fig.8: Front view of the Peltier connector

Section : 10 x 0.25 mm (2 x 3 x 0.25 for Peltier connections)