Hach 900 MAX manual Appendix D, Example, Calculate Q

Models: 900 MAX

1 160
Download 160 pages 21.13 Kb
Page 137
Image 137
Calculate Q.

Appendix D

Use the following examples to help you determine the value that you need to enter when programming for sampling based on counts.

Example 1

You want to collect 24 samples over a 24-hour period on a 4–20 mA flow proportional basis.

Average Flow Rate = 3.5 mgd (average flow rate over the 24-hour period) Maximum Flow Rate = 10 mgd

a = 1440 minutes

1.

Calculate Q.

 

 

Q = --Average------------------------Flow---------------Rate-------------

= 3.5----------mgd----------- = 0.35 mgd

 

Maximum Flow Rate

10 mgd

2.

Calculate t.

 

a 1440 min

t = --= ------------------------------ = 60 min/sample

n24 samples

3.Multiply Q x t x 10.

0.35mgd 60 min/sample 10 = 210

Therefore, the value entered for INTV = _ _ _ _ CNTS is 210.

Example 2

You want to collect 48 samples over a 24-hour period on a 4–20 mA flow proportional basis.

Average Flow Rate = 1.75 mgd (average flow rate over the 24-hour period) Maximum Flow Rate = 3 mgd

a = 1440 minutes

1.

Calculate Q.

 

 

Q = --Average------------------------Flow---------------Rate-------------

= 1.75--------------mgd----------= 0.583 mgd

 

Maximum Flow Rate

3 mgd

2.

Calculate t.

 

a 1440 min

t = --= ------------------------------ = 30 min/sample

n48 samples

3.Multiply Q x t x 10.

0.583 mgd 30 min/sample 10 = 174.9

Therefore, the value entered for INTV = _ _ _ _ CNTS is 175.

Example 3

You want to collect 96 samples over a 24-hour period on a 4–20 mA flow proportional basis.

Average Flow Rate = 0.52 mgd (average flow rate over the 24-hour period) Maximum Flow Rate = 2 mgd

a = 1440 minutes

1. Calculate Q.

Q = --Average------------------------Flow---------------Rate-------------

= 0.52--------------mgd----------= 0.26 mgd

Maximum Flow Rate

2.0 mgd

 

 

 

Page 135

8990apdx_pulsecounts.fm

How to Calculate Pulses/Counts

Page 137
Image 137
Hach 900 MAX manual Appendix D, Example, Calculate Q