Step 18 - Upset Sample

Section 3

17-D.Enable or Disable Continuous Mode using the CHANGE CHOICE key. If Continuous Mode is enabled the program will continuously run and switch bottle sets after every specified duration elapses, until manually stopped. If Continuous Mode is disabled, then the sampling halts when the last bottle set in the tray is full.

17-E.Press ACCEPT to continue to Upset Sample.

Step 18 - Upset Sample

Note: Unlike Setpoint Sampling, Upset Sampling can be enabled while the sampler is performing its regular sampling program. There must be more than one bottle in the sampler to perform Upset Sampling.

Upset Sampling analyzes the collected samples to determine when samples meet or exceed specified upset limits. The sampler will swing the distributor arm around to a sequestered set of bottles and grab an upset bottle.

The upset sample bottles are always the last bottles in the tray. For example, if there are 24 bottles in the tray, 4 first flush bottles and 4 upset bottles, then the bottles assignment is 1–4 first flush, 5–20 main program, and 21–24

upset samples.

Upset Samples are collected even when there are no more regular samples to be taken; when the status screen says Program Complete, but the bottom line of the display says Program Running.

Refer to the Quick Start Guides on page 103 for more setup information for individual channels.

18-A.Highlight Upset Sampling using the UP and DOWN keys on the Advanced Sampling Menu. Press SELECT to continue.

11:00 AM 21 - APR - 01

ADVANCED SAMPLING

SELECT

RETURN

START/STOP TIMES STORM WATER TIMED BOTTLE SETS

UPSET SAMPLING

VARIABLE INTERVALS VARIABLE VOLUME

18-B.Enable or Disable Upset Sampling using the CHANGE CHOICE key. Press ACCEPT to continue.

18-C.Press CHANGE CHOICE to select the desired channel to trigger from. Press ACCEPT to make the selection.

18-D.Press either HIGH CONDITION or LOW CONDITION.

Flow Rate of Change and Rainfall are signals that only increase in value and do not require Low Condition.

The external control signal (such as an external flow meter) must be configured in the external equipment for the desired setpoint.

 

11:00 AM 21 - APR - 01

UPSET SAMPLES

 

 

HIGH

 

UPSET CONDITION:

CONDITION

 

 

LOW

 

 

CONDITION

 

 

 

 

 

 

Page 58

 