d.Press ACCEPT to continue to Sample Distribution.

Section 3

b.Enter the Interval Between Samples. Press ACCEPT to continue.

11:00 AM 21 - APR - 01

SAMPLE COLLECTION

 

 

ACCEPT

INTERVAL:

00:00 (hrs:min)CLEAR

CANCEL

ENTRY

ENTER: 000:01—999.00 (hrs:min)

c.Select either Take First Sample Immediately or After The First Interval?

Note: When the program is started, the first sample is taken immediately upon pressing the start button or after the first interval has elapsed.

11:00 AM 21 - APR - 01

SAMPLE COLLECTION

ACCEPT

 

CHANGE

TAKE FIRST SAMPLE:

CHOICE

 

 

_______________

 

CANCEL

 

 

CHOICES: IMMEDIATE, AFTER 1st SAMPLE

d.Press ACCEPT to continue to Sample Distribution.

Flow Proportional Constant Volume, Variable Time (CVVT)

CVVT sampling is a flow-proportional method of sampling. CVVT samples are taken when a user-defined flow volume occurs in the flow stream. This occurs when the sampler is programmed to take a sample every time a specified volume of flow occurs in the flow stream. Specified flow volume intervals can vary with the flow rate of the stream, therefore varied time intervals can occur when constant (fixed) volume occurs.

Flow volume is determined internally, by the optional integral flow meter or by an external flow meter.

a.Press CHANGE CHOICE until Flow Proportional is displayed. Press

ACCEPT.

11:00 AM 21 - APR - 01

SAMPLE COLLECTION

 

 

 

CHANGE

ACCEPT

 

 

UNITS

 

SAMPLE COLLECTION:

 

 

 

FLOW-PROPORTIONAL

 

 

CANCEL

 

 

 

 

 

 

 

 

ENTER: 1.00 - 99999999

 

 

 

 

 

 

b.In the Flow Proportional menu, press CHANGE CHOICE until Constant Volume, Variable Time is displayed. Press ACCEPT.

 

11:00 AM 21 - APR - 01

FLOW PACING

 

 

 

 

 

CHANGE

 

 

ACCEPT

 

 

CHOICE

 

 

FLOW PACING MODE:

 

 

 

 

 

CONST VOL / VAR TIME

 

 

 

 

CANCEL

 

 

 

 

 

 

 

 

 

 

 

 

CHOICES: VAR T / CST VOL, CST T / VAR VOL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 41

8990startup.fm

Basic Programming Setup