Hach 900 MAX manual Section, Accept

Models: 900 MAX

1 160
Download 160 pages 21.13 Kb
Page 61
Image 61
Manual background

Section 3

18-E.Enter the desired high or low trigger point using the numeric keypad. Press ACCEPT to continue.

11:00 AM 21 - APR - 01

 

UPSET SAMPLES

ACCEPT

UPSET SAMPLING:

 

 

HIGH TRIGGER POINT:

 

CANCEL

00000 in.

CLEAR

 

ENTRY

 

 

 

 

 

 

(USE NUMERIC KEYPAD)

 

 

 

 

18-F.Enter the Deadband value or, if programming for Flow Rate Of Change or Rainfall, enter a time interval when the flow or rainfall change must take place (refer to Alarm Relays Programming on page 91).

11:00 AM 21 - APR - 01

UPSET SAMPLES

ACCEPT

UPSET SAMPLING

 

 

DEADBAND:

 

CANCEL

0.000

CLEAR

 

ENTRY

 

 

 

 

 

 

(USE NUMERIC KEYPAD)

 

 

 

 

18-G.Enter the number of bottles (out of the total set) to set aside for upset samples. These will be the last bottles in the tray.

11:00 AM 21 - APR - 01

UPSET SAMPLES

ACCEPT

UPSET SAMPLING:

 

 

NUMBER OF BOTTLES

 

CANCEL

2

CLEAR

 

ENTRY

 

 

 

 

 

 

ENTER 1 — 4

 

 

 

 

18-H.Select the method of distribution, Sample Per Bottle or Bottles Per Sample.

18-I.Enter the sample volume using the numeric keypad.

11:00 AM 21 - APR - 01

UPSET SAMPLES

ACCEPT

UPSET SAMPLING

 

 

SAMPLE VOLUME:

 

CANCEL

0000

CLEAR

 

ENTRY

 

 

 

 

 

 

ENTER 10 — 9999

 

 

 

 

18-J.Press ACCEPT to continue to Variable Intervals.

 

Page 59

8990startup.fm

Advanced Sampling

Page 61
Image 61