Section 3

a.In the Sample Collection menu, press CHANGE CHOICE until Flow Proportional is displayed. Press ACCEPT.

11:00 AM 21 - APR - 01

SAMPLE COLLECTION

 

 

 

CHANGE

ACCEPT

 

 

UNITS

 

SAMPLE COLLECTION:

 

 

 

FLOW-PROPORTIONAL

 

 

CANCEL

 

 

 

 

 

 

 

 

ENTER: 1.00 - 99999999

 

 

 

 

 

 

b.In the Flow Pacing menu, press CHANGE CHOICE until Const Time/Var Vol appears. Press ACCEPT.

11:00 AM 21 - APR - 01

 

FLOW PACING

 

 

 

 

 

 

CHANGE

 

ACCEPT

 

CHOICE

FLOW PACING MODE:

 

 

 

CONST TIME/ VAR VOL

 

 

 

CANCEL

 

 

 

 

 

 

 

 

CHOICES: VAR T / CST VOL, CST T / VAR VOL

c.In the Average Flow Rate menu use the numeric keypad to enter the known historical Average Flow Rate for a particular site.

11:00 AM 21 - APR - 01

CONST TIME/ VAR VOL

ACCEPT

AVERAGE FLOW RATE:

 

150.00 gph

CANCEL

CLEAR

ENTRY

 

 

 

 

ENTER: 1.00—99999999

 

 

d.In the Interval menu use the numeric keypad to enter the time interval between sample collections. Press ACCEPT.

11:00 AM 21 - APR - 01

CONST TIME/ VAR VOL

ACCEPT INTERVAL:

0:02 (hrs:min)

CLEAR

CANCEL

ENTRY

ENTER: 1.00—999:00 (hrs:min)

e.In the Total Volume Desired menu use the numeric keypad to enter the total volume of the sample. Press ACCEPT.

11:00 AM 21 - APR - 01

CONST TIME/ VAR VOL

ACCEPT

TOTAL VOLUME DESIRED:

 

1500 Ml

CANCEL

CLEAR

ENTRY

 

 

 

 

ENTER: 10—9999

 

 