Hach 900 MAX manual Appendix C, Page, Troubleshooting and Error Messages

Models: 900 MAX

1 160
Download 160 pages 21.13 Kb
Page 130
Image 130
Appendix C

Appendix C

Stormwater mode is not allowed when the sampler is

Tried to enable Storm Water with a single bottle.

configured with one bottle.

 

 

 

 

The program cannot start: timed rinse required.

Occurs if the liquid sensor is disabled and timed calibration

has not been performed for the rinse cycle.

 

 

 

 

The program cannot start: the primary device is not properly

If the operator goes partially through the flow meter setup

set up.

screens but stops before entering all necessary parameters.

 

 

 

 

 

If the liquid sensor is disabled and timed calibration has not

The program cannot start: volume calibration required.

been performed for the main sample, the first flush or the

 

 

upset sample volume.

 

 

The program setup needs flow to be logged at an

If an alarm, setpoint or stormwater uses flow rate-of-change

interval < or = the flow time limit.

as a trip point.

 

 

The program setup needs rain to be logged at an interval < or =

If an alarm, setpoint or stormwater uses rainfall as a

the rainfall time limit.

trip point.

 

 

 

 

 

When calibrating pH, process temperature, ORP, D.O., D.O.

 

 

temperature, conductivity, or conductivity temperature, the

The signal did not stabilize in time—calibration failed.

signal from the probe did not respond within the calibration

 

 

time limits. Replace buffer solutions (if applicable) or try

 

 

another probe.

 

 

 

The time entered is before the first logged value. Data will

 

 

be display from beginning.

When selecting the time/date for graphing and the selected

The time entered is after the last logged value. Data will be

beginning or ending time falls outside the range of data.

 

displayed from the end.

 

 

 

 

The value must be: xxx—yyy

The value entered was out of the legal range, when

changing any numeric parameter. The appropriate range is

(xxx is minimum, yyy is maximum).

always displayed on the bottom line on the entry screen.

 

 

 

 

 

This program stop is prior to the last program start.

When entering start/stop times. Start/stop times must be

This program start is prior to the last program stop.

entered in chronological order.

 

 

 

Timed bottle sets are not allowed unless distribution is

Tried to enable Timed Bottle Sets when bottles per sample

have not been specified (the bottles-per-sample is used as

bottles-per-sample mode.

the bottle set quantity).

 

 

 

 

Timed bottle sets are not allowed when the sampler is

Tried to enable Timed Bottle Sets with a single bottle.

configured with one bottle.

 

 

 

 

Times cannot be more than 24 hours apart.

When selecting a start and stop time for a graph display the

limit (due to screen resolution) is 24 hours.

 

 

 

 

 

Too few bottles—review timed bottle sets, upset sampling, &

Occurs if there are not enough bottles to perform all the

various functions. Either install more bottles or decrease

stormwater. Minimum needed:

number of programmed functions.

 

 

 

 

 

Too many alarms and setpoints have been requested.

Occurs if more than 20 alarm and setpoint conditions are

set. Storm Water start conditions use one or two setpoints.

 

 

 

 

Too many samples-per-bottle for the bottle volume:

Occurs when the sample volume is checked, if the

The sample volume is too large for the bottle volume:

requested sample volume is too big to fit in the bottle.

 

 

Upset sampling is not allowed when the sampler is configured

Tried to enable Upset Sampling with a single bottle.

with one bottle.

 

 

 

 

Page 128

 

Troubleshooting and Error Messages

8990apdx_tbl.fm

Page 130
Image 130
Hach 900 MAX manual Appendix C, Page, Troubleshooting and Error Messages