Hach 900 MAX manual 5.2.2pH Probe Programming, 5.2.3pH Probe Calibration, Section

Models: 900 MAX

1 160
Download 160 pages 21.13 Kb
Page 73
Image 73
5.2.2pH Probe Programming

Section 5

Figure 22 pH Probe Wiring to Junction Box (grounded)

GND

REF

RTD

RTD

 

 

 

 

Red 5.2.3pH Probe Calibration

Black

pH

Glass

Figure 23 pH Probe Wiring to Junction Box (un-grounded)

GND

REF

RTD

RTD

 

 

 

 

 

 

 

 

Red

Yellow

Green

Clear

Yellow

Green

pH

Clear

Glass

5.2.2pH Probe Programming

1.From the Main Menu, select OPTIONS>ADVANCED OPTIONS>DATALOG.

2.Highlight Select Inputs using the UP and DOWN keys and press SELECT.

3.Highlight pH using the UP and DOWN keys, then press SELECT.

4.Press the CHANGE CHOICE key to cycle between Logged and Not Logged, then press ACCEPT.

5.Enter a logging interval, then press ACCEPT. Valid logging intervals are shown on the status bar along the bottom edge of the display.

6.Select another channel to configure or press RETURN to back up one step. Press MAIN MENU to return to the Main Menu display.

5.2.3pH Probe Calibration

Calibrate the pH probe after the pH probe is connected and programmed. Calibrating the pH probe requires a thermometer and any two of the following buffer solutions: 4, 7, or 10 pH.

The pH probe is an application sensitive device. When used in harsh environments, the accuracy and life expectancy of pH probes can decrease.

 

Page 71

8990options.fm

pH Probe

Page 73
Image 73
Hach 900 MAX manual 5.2.2pH Probe Programming, 5.2.3pH Probe Calibration, Section