Hach 900 MAX Appendix B Options Menu Features, Setting the Time and Date, Volume Calibration

Models: 900 MAX

1 160
Download 160 pages 21.13 Kb
Page 118
Image 118
Appendix B

Appendix B

Options Menu Features

1.From the Main Menu, select OPTIONS.

11:00 AM 21 - APR - 01

OPTION MENU

 

VOLUME

TIME / DATE

CALIBRATION

ADVANCED

LEVEL

OPTIONS

ADJUST

READY TO START

 

 

 

 

Optional Programs

Time and Date for internal real-time clock

Calibrate the sample volume

Advanced Features

Adjust level readings of optional integral flow meter

Setting the Time and Date

1.From the Main Menu, select OPTIONS>TIME/DATE

11:00 AM 21 - APR - 01

TIME / DATE

 

ACCEPT

 

CHANGE

 

 

AM / PM

 

 

 

 

 

_ _: _ _ AM _ _ -APR- _ _

 

 

CLEAR

MODE: 12-HR FORMAT

CHANGE

 

ENTRY

MONTH

 

 

 

 

 

 

 

 

USE +/- KEY TO CHANGE 12/24 HR FORMAT

 

 

 

 

 

2.Adjust the real time clock and date setting, using CHANGE AM/PM and

CHANGE MONTH keys.

Volume Calibration

From the Main Menu, select OPTIONS> VOLUME CALIBRATION.

The liquid sensor on the sampler offers unparalleled sample accuracy and repeatability. For most site conditions, no volume calibration is necessary, and the volume collected is accurate within 10% of the requested volume. To improve volume accuracy to 5%, or when sample volume accuracy is adversely affected by unusual site conditions (such as the sample source located higher than the sampler pump), or when the liquid sensor is disabled, then volume calibration should be performed.

Sensor Disabled

When the sensor is disabled, you must calibrate the sample volume(s). If more than one volume is programmed (such as when using storm water or upset sampling), you must calibrate each volume independently (see chart below).

Page 116

Programming Features

8990apdx_pfeatures.fm

Page 118
Image 118
Hach 900 MAX manual Appendix B Options Menu Features, Setting the Time and Date, Volume Calibration, Sensor Disabled