9600 Compressor Description

CTI-CRYOGENICS

HELIX TECHNOLOGY CORPORATION

Cooling Water

The water used to cool the Compressor must meet the specifications shown in Table 1-3 for proper system operation.

Table 1-3: Cooling Water Specifications

Parameter

Value

 

 

Maximum Inlet Temperature

90°F (32°C)

 

 

Minimum Inlet Temperature

50°F (10°C)

 

 

Flow Rate

2.75 ±1.25 gpm (10.4 ± 4.7 lpm)

 

 

Pressure Drop (inlet-to-outlet)

See Figure 1-3

 

 

Maximum Inlet Pressure

100 psi (6.9 bars)

 

 

Alkalinity

6.0 - 8.0 pH

 

 

Calcium Carbonate

< 75 ppm

 

 

NOTE: Water conditioning may be required for applications not meeting these requirements.