Appendix C - Troubleshooting Procedures

CTI-CRYOGENICS

HELIX TECHNOLOGY CORPORATION

Table C-1: Compressor Troubleshooting Procedures

Problem

 

Possible Cause

 

Corrective Action

 

 

 

 

 

1) System circuit breaker

1)

Incorrect phasing of input

1)

Check phasing of input

(CB1) trips immediately

 

power.

 

power. Refer to Phase

to the OFF (0) position

 

 

 

Check in Section 3.

when switched to the ON

 

 

 

 

(1) position.

 

 

 

 

 

 

 

 

 

2) System (CB1) and Con-

1)

No power coming from

1)

Check source fuses, cir-

trol Circuit (CB2) circuit

 

source.

 

cuit breakers, and wiring

breakers remain in the

 

 

 

associated with the

ON (1) position when

 

 

 

power source. Repair as

switched ON but the

 

 

 

needed.

compressor will not run.

 

 

 

 

 

2)

Insufficient power

2)

Verify adequate phase-to-

 

 

 

 

phase input voltage.

 

 

 

 

Refer to Table 1-2.

 

3)

Remote control jumper

3)

Check to insure that

 

 

plug not in place. This will

 

remote jumper plug is

 

 

apply only if remote cir-

 

fully seated. See Figure

 

 

cuit is not being used.

 

1-4 for location. Refer to

 

 

 

 

Compressor Remote

 

 

 

 

Connector in Section 1

 

 

 

 

for more information.

 

4)

Improperly wired external

4)

Verify correct installation

 

 

remote control circuit.

 

of remote control feature.

 

 

NOTE: Only applies if

 

Refer to Table 1-6.

 

 

remote control feature is

 

 

 

 

being used.

 

 

 

 

 

 

 

C-2

P/N 8040540

Page 55
Image 55
Olicom 9600 manual Table C-1 Compressor Troubleshooting Procedures, Compressor Remote