Installation
CTI-CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Electrical Connections
The following procedures provide information for making all three phase (180 - 250 VAC) electrical connections to the Compressor.
Power Cable Preparation
WARNING
Follow all local high voltage safety precautions when performing this procedure to reduce the possibility of electrical shock. Make sure all electrical power is OFF before proceeding with this procedure.
CAUTION
The cable used for making the Compressor power cable must be 10 gauge, 3 conductor cable with ground rated at 600 VAC.
1.Cut a 10 AWG (6.00 mm2), 3 conductor cable with ground to an appropriate length.
2.Strip the cable jacket back 4 in. (101.6 mm).
3.Strip the insulation back 3/8 in. (9.3 mm) on each individual conductor.
4.Install a #10 ring tongue terminal on the end of each conductor using the appropriate size double crimping tool.
5.Remove the rear panel as shown in Figure
6.Remove the circuit breaker terminal cover as shown in Figure
7.Install the cable into the Compressor through the cable strain relief.
8.Remove the
NOTE: Use a slotted screwdriver which is capable of holding a screw when performing steps 9 and 10.
9.Remove the screws from the Compressor circuit breaker terminals X, Y, and Z as shown in Figure
P/N 8040540 |