47 - iRIS 220 / iRIS 320, V1.19 User Guide

Sensor Settings Screen 5/5 - Alarm 2 (Level 5)

Sensor Alarm screen 2 shows the same information as Alarm screen 1, but for Alarm #2.

ALARM#2 5/5

/6.1000

\ 6.2500 ;(min) 00000

+/-

Move forwards/backwards through sensor setting screens.

Alt Enter

Move up to Sensor Status screen

Sensor Calibration Screen (Level 5)

The Sensor Calibration screen is used for entering a calibration offset. The top value on the display is the current sensor reading. The bottom value on the display is the actual reading as determined from a gauge board or similar physical reference. The calibration offset is automatically calculated from these two values.

CALIBRATION 11.6000

+0.8000 = 12.4000

+/-

Increment / decrement actual value

Enter

Multiply actual value by factor of 10

Alt Enter

Move to Calibration Acceptance screen

Sensor Calibration Acceptance Screen (Level 5)

The Sensor Calibration Acceptance screen is used to accept or decline the sensor calibration.

ACCEPT?

>No

Yes

+/-

Move down / up through available options

Enter

Accept current selection

Sensor Data Screen (Level 5)

The Sensor Data screen is used to view the logged sample data for a sensor. The data pointer value for the top sample is displayed on the top right hand side.

DATA 108 1375 13:00 1372 13:30 1371 14:00

+/-

Move forwards/backwards through sample values.

Alt Enter

Move up to Sensor Status screen

Alt +/-

Pan left/right to view the rest of the sample time & date information.

5.2.7 Totaliser Related Screens

The Totaliser screens show yesterday's total, the daily (since 00:00:00 today) and running totals for the two pulse input counters.

Total Screen 1/2 (Level 3)

TOTAL

½

Yst

134.6

Day

17.8

Run

5432.4

 

 

+/-

Move forwards/backwards through sensor total screens

Enter

Drill down to Total Reset screen

Alt Enter

Move up to Main Menu screen

Page 49