5 Managing Patients

Care Groups

Patient Mismatch - If Both Patient Data Sets Are Correct

A patient mismatch where both sets of patient data are correct might occur if you admit a new patient at the monitor (or Information Center) before the patient arrives at your unit and then connect the MMS that was used during the patient transport to the monitor.

Select Same Patient if the patient information is different, but you are sure it is the same patient. This merges the demographics and updates them in the Information Center, monitor, and MMS, according to this table. Be aware that your monitor may be configured to merge trend data from the MMS and the monitor, and to upload measurement settings from the MMS to the monitor.

Patient Information

This information is taken from...

 

 

Patient name

the monitor, if the patient was admitted there. For centrally-admitted

 

patients, this information is taken from the Information Center.

MRN

 

 

 

Screen Notes

 

 

 

Patient Category

the Multi-Measurement Server. if connected, otherwise the data is

 

taken from the monitor.

Date of Birth

 

 

 

Height

 

 

 

Weight

 

 

 

Gender

 

 

 

Paced Status

Paced status is always set to Yes where there is a conflict in patient

 

information.

 

 

Trend data

if there is newer trend data stored in the MMS, it is uploaded to the

 

monitor.

 

 

Automatically Resolving Patient Mismatch

Your monitor can be configured to automatically resolve mismatches in one of two ways.

continue using the patient in the MMS, and delete the old data in the monitor. This is suitable for transport monitors.

continue with the patient in the monitor, and delete the data in the MMS.

Care Groups

If your monitor is connected to an Information Center, you can group bedside monitors into Care Groups. This lets you:

view information on the monitor screen from another bed in the same or in a different Care Group.

be notified of yellow or red alarm conditions at the other beds in the Care Group.

see the alarm status of all the beds in the Care Group on each monitor screen.

There are two main types of Care Groups:

standard care group (Bed-based Care Group) - up to 12 patients monitored by up to 4 Information Centers

unit group (Unit-based Care Group) - for a complete unit with up to 64 patients monitored by up to 4 Information Centers

84

Page 104
Image 104
Philips MP40/50 Care Groups, Automatically Resolving Patient Mismatch, Patient Information This information is taken from

MP60/70/80/90, MP40/50, MP20/30 specifications

Philips MP20/30, MP40/50, and MP60/70/80/90 are advanced patient monitoring systems designed to enhance clinical workflows and patient care in diverse healthcare settings. These devices combine cutting-edge technologies and user-friendly interfaces, making them essential tools for hospitals and healthcare providers.

The MP20 and MP30 are notable for their compact design, making them ideal for environments where space is limited. They come equipped with a variety of monitoring parameters, including ECG, SpO2, and non-invasive blood pressure, providing a comprehensive overview of patient health. These models are particularly suited for use in general wards or recovery areas, offering reliability and ease of use.

Moving to the MP40 and MP50, these monitors incorporate more advanced functionalities, including multi-gas monitoring, which is vital for anesthetic applications. They feature a robust data storage capacity, allowing for extensive patient history tracking and improved decision-making capabilities. The intuitive touchscreen interfaces facilitate quick access to essential information, enabling healthcare professionals to respond efficiently to changes in a patient's condition.

The MP60, MP70, MP80, and MP90 series represent the pinnacle of Philips' patient monitoring technology. These monitors are equipped with advanced algorithms that enhance the precision of vital sign measurements, including advanced ECG analysis capabilities that assist in detecting arrhythmias. The MP80 and MP90 models further integrate advanced monitoring technologies such as 3D visualization for hemodynamics and deep learning algorithms for early warning systems.

Connectivity is a key feature across the Philips monitoring range, with many models offering seamless integration with hospital information systems and electronic health records. This connectivity ensures that patient data is easily accessible to clinicians, promoting informed decision-making. Moreover, alarm management features are designed to minimize false alarms, significantly improving the staff's response time by focusing on critical alerts.

In summary, Philips MP20/30, MP40/50, MP60/70/80/90 monitors deliver exceptional monitoring capabilities, combining user-friendly design with advanced technology. Their varied features allow them to cater to different clinical needs, enhancing patient care and supporting healthcare professionals in their critical roles. With these monitors, healthcare facilities can achieve a new standard in patient monitoring, ensuring quality and safety in patient management.