4 Patient Alarms and INOPs

Technical Alarm Messages (INOPs)

INOP Message, Indication

Source

What to do

 

 

 

Ttymp INOPS

TEMP

See <Temp label> INOPs (under Temp)

Tven INOPS

TEMP

See <Temp label> INOPs (under Temp).

Tvesic INOPS

TEMP

See <Temp label> INOPs (under Temp)

UAP INOPS

PRESS

See <Pressure label> INOPS (under Pressure).

Unsupported LAN

Monitor

There is a problem with the communication to the network

INOP tone

 

and central monitoring is currently not possible. Check the

 

connection.If the INOP persists, switch off the monitor and

 

 

 

 

contact your service personnel.

User I/F Malfunct.

Monitor

Perform a visual and functional check of all the monitor input

INOP tone.

 

devices. Contact your service personnel.

UVP INOPS

PRESS

See <Pressure label> INOPS (under Pressure).

V LEAD OFF

ECG

The V electrode (IEC: C electrode) has become detached from

Numeric is replaced by -?- for 10

 

the patient or the lead set has been changed. Reattach the

seconds; INOP tone.

 

electrode or select New Lead Setup in the Setup ECG

 

 

menu to confirm the new lead set.

<VueLink option> CHK

VueLink

No cable or the wrong cable connected to the VueLink module,

CABLE

 

or incorrect device selected. Silencing this INOP switches the

INOP tone.

 

measurement off.

 

 

VueLink INOP abbreviations may differ slightly depending on

 

 

the device category.

<VueLink option> CHK

VueLink

The wrong external device has been selected on the VueLink

CONF.

 

module, or the external device has not been correctly setup, or

INOP tone.

 

the wrong cable has been used to connect the device to the

 

 

VueLink module.

 

 

VueLink INOP abbreviations may differ slightly depending on

 

 

the device category.

<VueLink option> CHECK

VueLink

No information was received from the external device. The

SETUP

 

device may be switched off or disconnected.

INOP tone.

 

VueLink INOP abbreviations may differ slightly depending on

 

 

the device category.

VueLnk EQUIP MALF

VueLink

Malfunction in the VueLink module. If this message appears

INOP tone.

 

repeatedly, the module must be replaced. Contact your service

 

 

personnel.

 

 

VueLink INOP abbreviations may differ slightly depending on

 

 

the device category.

VueLnk NO CONFIG

VueLink

The VueLink module has not been configured during

INOP tone.

 

installation. The installation process should be completed by

 

 

either your biomedical engineering department or the Philips

 

 

service engineer.

 

 

VueLink INOP abbreviations may differ slightly depending on

 

 

the device category.

VueLnk UNPLUGGED

VueLink

The VueLink module has been unplugged from the rack, or the

INOP tone.

 

whole rack has been disconnected. The measurement

 

 

automatically disappears from the display. Silencing this INOP

 

 

switches off the measurement.

 

 

VueLink INOP abbreviations may differ slightly depending on

 

 

the device category.

 

 

 

76

Page 96
Image 96
Philips MP60/70/80/90, MP20/30, MP40/50 manual Cable, Conf, Setup

MP60/70/80/90, MP40/50, MP20/30 specifications

Philips MP20/30, MP40/50, and MP60/70/80/90 are advanced patient monitoring systems designed to enhance clinical workflows and patient care in diverse healthcare settings. These devices combine cutting-edge technologies and user-friendly interfaces, making them essential tools for hospitals and healthcare providers.

The MP20 and MP30 are notable for their compact design, making them ideal for environments where space is limited. They come equipped with a variety of monitoring parameters, including ECG, SpO2, and non-invasive blood pressure, providing a comprehensive overview of patient health. These models are particularly suited for use in general wards or recovery areas, offering reliability and ease of use.

Moving to the MP40 and MP50, these monitors incorporate more advanced functionalities, including multi-gas monitoring, which is vital for anesthetic applications. They feature a robust data storage capacity, allowing for extensive patient history tracking and improved decision-making capabilities. The intuitive touchscreen interfaces facilitate quick access to essential information, enabling healthcare professionals to respond efficiently to changes in a patient's condition.

The MP60, MP70, MP80, and MP90 series represent the pinnacle of Philips' patient monitoring technology. These monitors are equipped with advanced algorithms that enhance the precision of vital sign measurements, including advanced ECG analysis capabilities that assist in detecting arrhythmias. The MP80 and MP90 models further integrate advanced monitoring technologies such as 3D visualization for hemodynamics and deep learning algorithms for early warning systems.

Connectivity is a key feature across the Philips monitoring range, with many models offering seamless integration with hospital information systems and electronic health records. This connectivity ensures that patient data is easily accessible to clinicians, promoting informed decision-making. Moreover, alarm management features are designed to minimize false alarms, significantly improving the staff's response time by focusing on critical alerts.

In summary, Philips MP20/30, MP40/50, MP60/70/80/90 monitors deliver exceptional monitoring capabilities, combining user-friendly design with advanced technology. Their varied features allow them to cater to different clinical needs, enhancing patient care and supporting healthcare professionals in their critical roles. With these monitors, healthcare facilities can achieve a new standard in patient monitoring, ensuring quality and safety in patient management.