Technical Alarm Messages (INOPs) 4 Patient Alarms and INOPs
73
<SpO2 LABEL> UPGRADE
Label is displayed with a -?-, numeric is
unavailable
SpO2The SpO2 measurement is currently in UPGRADE mode.
Monitoring is not possible in this mode.
Sp - vO2 CHK SOURCES
Numeric is replaced by -?-
Sp - vO2Not all measurements or values required to perform the
calculation are available. Check measurement sources.
Sp - vO2 CHK UNITS
Numeric is replaced by -?-
Sp - vO2The monitor has detected a conflict in the units used for this
calculation. Check the unit settings.
SvO2 CAL FAILED
SvO2 numeric is displayed with ?
SvO2The calibration failed. Check the catheter-to-Optical-Module
connection. Manually restart the calibration. Try another
catheter and Optical Module. If the catheter is already inserted,
perform an in-vivo calibration.
SvO2 CAL MODE
SvO2 numeric is replaced by -?-
SvO2Pre-insertion calibration is complete, but the catheter tip is still
inside the optical reference. The catheter is now ready for
insertion.
SvO2 CAL REQUIRED
SvO2 numeric is replaced by -?-.
INOP tone may sound
SvO2There is no valid calibration data in the Optical Module.
Perform either a pre-insertion or an in-vivo calibration.
SvO2 CONFIGURATION
SvO2 numeric is replaced by -?-.
INOP tone
SvO2The Optical Module has been configured to SaO2 Mode. Use
Change to SvO2 in the Setup S vO2 menu to
reconfigure to SvO2 Mode.
SvO2 CONNCT OPTMOD
SvO2 numeric is replaced by -?-.
INOP tone
SvO2The Optical Module was disconnected during data storage.
Reconnect the Optical Module for at least 20 seconds.
SvO2 EQUIP MALF
SvO2 numeric is replaced by -?-.
INOP tone
SvO2The SvO2 Module or Optical Module is faulty. Unplug and
replug the Optical Module and SvO2 module. Exchange the
modules. If the INOP persists, contact your service personnel.
SvO2 IN-VIVO CALIB
SvO2 numeric is replaced by -?-.
SvO2The in-vivo calibration is not yet complete. Lab values must be
stored to the Optical Module to complete the calibration.
Either continue with the next steps of the current calibration or
recall the previous calibration.
SvO2 LIGHT INTENS
SvO2 numeric is replaced by -?- or
numeric is displayed with ?
INOP tone with -?- display
SvO2The intensity changed considerably since the last light intensity
calibration. This may indicate that the catheter tip is positioned
against a blood vessel wall or that there is low blood flow.
Reposition the catheter and perform a light intensity
calibration.
SvO2 LOW LIGHT
SvO2 numeric is replaced by -?- or
numeric is displayed with ?
INOP tone may sound
SvO2The optical signal levels are too low. Check that the catheter is
either in the optical reference or inserted into the patient.
Check the catheter-to-Optical Module connection. If INOP
persists, try another catheter and Optical Module.
SvO2 NO OPTMOD
SvO2 numeric is replaced by -?-.
INOP tone
SvO2Connect the Optical Module. If the INOP persists, try another
Optical Module. Silencing this INOP switches the
measurement off.
SvO2 OPTMOD DEFECT SvO2The Optical Module memory is faulty, and calibration data
cannot be stored for transport or during power failure. If this
feature is needed, use another Optical Module.
SvO2 OPTMOD WARMUP
SvO2 numeric is displayed with ?
SvO2The Optical Module has not yet reached the operating
temperature. Wait a few minutes until warm-up is finished.
SvO2 PRE-INS CALIB
SvO2 numeric is replaced by -?-
INOP tone
SvO2The pre-insertion calibration is running. This typically takes
one minute. During this time SvO2 alarms are switched off.
Wait until the calibration is complete.
INOP Message, Indication Source What to do