17 Monitoring SvO2

Preparing to Monitor SvO2

Preparing to Monitor SvO2

In addition to an SvO2 module, you need a Hospira OptiCath catheter, and 50131 Optical Module. Use only the Hospira accessories listed in the Accessories section.

1optical module

2balloon

 

inflation

4

 

 

 

stopcock

 

3

Hospira fiber

 

 

optic catheter

5

4optical reference

5enter setup/ calibration

3

1

2

Connect the optical module (Hospira 50131) to the SvO2 module. Allow the optical module to warm up before you perform a calibration. Although the warm up message disappears from the screen after one minute, Hospira recommends letting the optical module warm up for 15 minutes for best accuracy. Please refer to the instructions for the optical module.

To avoid false alarms during the pre-insertion calibration and insertion of the catheter into the patient, the monitor automatically suspends alarms during the pre-insertion calibration, for up to three minutes after you remove the catheter tip from the optical reference. After light intensity calibration, or after three minutes (whichever comes first), the monitor returns to the alarm state it was in prior to pre- insertion calibration.

Carrying out a Pre-insertion Calibration

WARNING It is strongly recommended to carry out a pre-insertion calibration prior to all insertions. If this is not possible, you must perform an in-vivo calibration after insertion.

Refer to the instructions for use that accompany the catheter. Do not use the catheter if the packaging is damaged. If you have to disconnect the monitor from the patient (for example, when transferring the patient from one location to another), you must disconnect at the SvO2 module. The catheter should remain in the optical module, otherwise you need to recalibrate.

1Remove outer wrapping from catheter tray to uncover optical connector.

2Place the optical module on the catheter tray in the space provided and open the lid.

204

Page 224
Image 224
Philips MP40/50, MP60/70/80/90, MP20/30 manual Preparing to Monitor SvO2, Carrying out a Pre-insertion Calibration

MP60/70/80/90, MP40/50, MP20/30 specifications

Philips MP20/30, MP40/50, and MP60/70/80/90 are advanced patient monitoring systems designed to enhance clinical workflows and patient care in diverse healthcare settings. These devices combine cutting-edge technologies and user-friendly interfaces, making them essential tools for hospitals and healthcare providers.

The MP20 and MP30 are notable for their compact design, making them ideal for environments where space is limited. They come equipped with a variety of monitoring parameters, including ECG, SpO2, and non-invasive blood pressure, providing a comprehensive overview of patient health. These models are particularly suited for use in general wards or recovery areas, offering reliability and ease of use.

Moving to the MP40 and MP50, these monitors incorporate more advanced functionalities, including multi-gas monitoring, which is vital for anesthetic applications. They feature a robust data storage capacity, allowing for extensive patient history tracking and improved decision-making capabilities. The intuitive touchscreen interfaces facilitate quick access to essential information, enabling healthcare professionals to respond efficiently to changes in a patient's condition.

The MP60, MP70, MP80, and MP90 series represent the pinnacle of Philips' patient monitoring technology. These monitors are equipped with advanced algorithms that enhance the precision of vital sign measurements, including advanced ECG analysis capabilities that assist in detecting arrhythmias. The MP80 and MP90 models further integrate advanced monitoring technologies such as 3D visualization for hemodynamics and deep learning algorithms for early warning systems.

Connectivity is a key feature across the Philips monitoring range, with many models offering seamless integration with hospital information systems and electronic health records. This connectivity ensures that patient data is easily accessible to clinicians, promoting informed decision-making. Moreover, alarm management features are designed to minimize false alarms, significantly improving the staff's response time by focusing on critical alerts.

In summary, Philips MP20/30, MP40/50, MP60/70/80/90 monitors deliver exceptional monitoring capabilities, combining user-friendly design with advanced technology. Their varied features allow them to cater to different clinical needs, enhancing patient care and supporting healthcare professionals in their critical roles. With these monitors, healthcare facilities can achieve a new standard in patient monitoring, ensuring quality and safety in patient management.