Technical Alarm Messages (INOPs)

4 Patient Alarms and INOPs

INOP Message, Indication

Source

What to do

 

 

 

CCO PRESS INVALID

C.O.

The arterial pressure selected for pulse contour calculation for

at Information Center

 

CCO is available but currently invalid. Make sure the pressure

 

 

transducer is connected and the zero calibration is valid.

CCO PRESS OVERRANG

C.O.

The mean value of the arterial pressure values used for pulse

numeric is replaced by -?-

 

contour calculation for CCO is below 0 mmHg or above 300

INOP tone

 

mmHg.

CCO PULSE OVERRANG

C.O.

The pulse rate of the pressure used for pulse contour

numeric is replaced by -?-

 

calculation for CCO is below 30 bpm or above 240 bpm.

INOP tone

 

 

CCO/Tbl NO TRANSD

C.O.

No transducer attached to the module or catheter

Numeric is replaced by -?-

 

disconnected.

INOP tone

 

 

CCO RECALIBRATE

C.O.

The most recent CCO or CCI calibration was made over 8

numeric is replaced by -?-

 

hours ago. You should recalibrate CCO or CCI with

 

 

transpulmonary C.O. measurements at least every 8 hours or

 

 

when the hemodynamic condition of the patient has changed.

Charge BATT1/BATT2 now

Batteries

Battery must be charged. Connect the monitor to mains power

INOP tone

 

or exchange the battery.

CHARGER MALFUNCT

Batteries

There is a problem with the battery charger in the monitor.

INOP tone, battery LED may flash

 

Connect the monitor to mains power and contact your service

 

personnel.

 

 

Check Alarm Lamps

Monitor

Perform a visual check of the alarm lamp to establish whether

INOP tone.

 

there is a problem. Contact your service personnel to check the

 

 

internal connections to the alarm lamps.

CHECK BATT TEMP

Battery

The temperature of one or both batteries is too high. Check

INOP tone

 

that ventilation openings are not blocked and monitor is not

 

 

exposed to heat.

Check Flex Texts

Monitor

Check the names of the monitor menus, for example the labels

INOP tone

 

for screens, profiles, event or trend group names, before you

 

 

resume monitoring. If they are unexpected, there may be a

 

 

problem with the monitor software. Contact your service

 

 

personnel.

CheckInternVoltage

Monitor

Potential problem with alarm lamps, display or interfaces

at Information Center

 

detected. Contact your service personnel. This INOP will

 

 

appear on the monitor as Check Monitor Func.

Check Keyboard

Monitor

Perform a visual and functional check of the keyboard. Contact

INOP tone

 

your service personnel.

 

 

Check Main Board 2

Monitor

There is a problem with the second main board in the monitor.

INOP tone.

 

Contact your service personnel.

Check MCC

Monitor

The monitor cannot communicate with the D80 Intelligent

 

 

Display. Check the MSL coupling cable. The end with the grey

 

 

connector must be connected to the Intelligent Display.

Check Monitor Func

Monitor

Potential problem with alarm lamps, display or interfaces

INOP tone.

 

detected. Contact your service personnel. This INOP may

 

 

appear on the Information Center as

 

 

CheckInternVoltage.

Check Monitor Temp

Monitor

The temperature inside the monitor is too high. Check that the

INOP tone

 

monitor ventilation is not obstructed. If the situation

 

 

continues, contact your service personnel.

 

 

 

63

Page 83
Image 83
Philips MP40/50, MP20/30 manual CCO Press Invalid, CCO Press Overrang, CCO Pulse Overrang, CCO Recalibrate, Charger Malfunct

MP60/70/80/90, MP40/50, MP20/30 specifications

Philips MP20/30, MP40/50, and MP60/70/80/90 are advanced patient monitoring systems designed to enhance clinical workflows and patient care in diverse healthcare settings. These devices combine cutting-edge technologies and user-friendly interfaces, making them essential tools for hospitals and healthcare providers.

The MP20 and MP30 are notable for their compact design, making them ideal for environments where space is limited. They come equipped with a variety of monitoring parameters, including ECG, SpO2, and non-invasive blood pressure, providing a comprehensive overview of patient health. These models are particularly suited for use in general wards or recovery areas, offering reliability and ease of use.

Moving to the MP40 and MP50, these monitors incorporate more advanced functionalities, including multi-gas monitoring, which is vital for anesthetic applications. They feature a robust data storage capacity, allowing for extensive patient history tracking and improved decision-making capabilities. The intuitive touchscreen interfaces facilitate quick access to essential information, enabling healthcare professionals to respond efficiently to changes in a patient's condition.

The MP60, MP70, MP80, and MP90 series represent the pinnacle of Philips' patient monitoring technology. These monitors are equipped with advanced algorithms that enhance the precision of vital sign measurements, including advanced ECG analysis capabilities that assist in detecting arrhythmias. The MP80 and MP90 models further integrate advanced monitoring technologies such as 3D visualization for hemodynamics and deep learning algorithms for early warning systems.

Connectivity is a key feature across the Philips monitoring range, with many models offering seamless integration with hospital information systems and electronic health records. This connectivity ensures that patient data is easily accessible to clinicians, promoting informed decision-making. Moreover, alarm management features are designed to minimize false alarms, significantly improving the staff's response time by focusing on critical alerts.

In summary, Philips MP20/30, MP40/50, MP60/70/80/90 monitors deliver exceptional monitoring capabilities, combining user-friendly design with advanced technology. Their varied features allow them to cater to different clinical needs, enhancing patient care and supporting healthcare professionals in their critical roles. With these monitors, healthcare facilities can achieve a new standard in patient monitoring, ensuring quality and safety in patient management.