Measurement Specifications 34 Installation and Specifications
361
Cardiac Output / Continuous Cardiac Output
Negative Inspiratory
Pressure numeric
Range -20 to 0 cmH2O
Resolution 1 cmH2O
Mean Airway Pressure Range -20 to +120 cmH2O
Resolution 1 cmH2O
Plateau Pressure
numeric
Range 0 to +120 cmH2O
Resolution 1 cmH2O
Positive End-
Expiratory Pressure
numeric (PEEP)
Range 0 to 99 cmH2O
Resolution 1 cmH2O
Barometric Pressure
numeric (PB)
Range 400 to 800 mmHg
Resolution 1 mmHg
Peak Inspiratory Flow
numeric (PIF)
Range 0.1 to 25 l/min (neonatal)
1 to 100 l/min (pediatric)
2 to 180 l/min (adult)
Resolution 0.1 l/min (neonatal)
1 l/min (pediatric/adult)
Peak Expiratory Flow
numeric (PEF)
Range 0.1 to 25 l/min (neonatal)
1 to 100 l/min (pediatric)
2 to 180 l/min (adult)
Resolution 0.1 l/min (neonatal)
1 l/min (pediatric/adult)
Inspiration to
Expiration ratio
numeric (I:E)
Range 1: 0.2 ... 9.9
Resolution 0.1
Spirometry Performance Specifications

C.O./CCO Performance Specifications

C.O. (right heart) Range 0.1 to 20.0 l/min
Accuracy Instrument Specification (electrical):+ 3% or 0.1 l/min
System Specification:+ 5%.or 0.2 l/min
Repeatability Instrument Specification (electrical):+ 2% or 0.1 l/min
System Specification:+ 3%.or 0.1 l/min
C.O.
(transpulmonary)
Range 0.1 to 25.0 l/min
Accuracy Instrument Specification (electrical):+ 4% or 0.15 l/min
System Specification:+ 5% or 0.2 l/min
Repeatability Instrument Specification (electrical):+ 2% or 0.1 l/min
System Specification:+ 3% or 0.1 l/min
EVLW
not available in USA
Range 10 to 5000 ml
Standard Deviation 10% or 1 ml/kg