Measurement Specifications34 Installation and Specifications

tcGas Performance Specifications

tcpCO2 response time, including transducer

< 50 s

 

 

 

Warm-up Time

 

< 3 minutes

 

 

 

 

 

 

 

 

Site Timer

 

0.5, 1, 1.5, 2, 2.5, 3, 4, 5, 6, 7, or 8 hours. Change Site alarm

 

 

when site time is expired and configurable automatic heating

 

 

switch-off

 

 

 

 

 

 

 

 

Transducer Heating

 

Available Temperatures -

 

 

 

 

37.0°C, 41.0 to 45°C in steps of 0.5°C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

tcGas Alarm Specifications

Range

Adjustment

Alarm Delay

 

 

 

 

 

 

tcpO2

10 to 745 mmHg

10 to 30 mmHg: 1 mmHg

10 seconds after the

 

1.0 to 99.5 kPa

1.0 to 4.0 kPa: 0.1 kPa

value exceeds the set

 

 

 

32 to 100 mmHg: 2 mmHg

limit range.

 

 

 

 

 

 

 

 

4.2 to 13 kPa: 0.2 kPa

 

 

 

 

 

105 to 745mmHg: 5 mmHg

 

 

 

 

 

13.5 to 99.5 kPa: 0.5 kPa

 

 

tcpCO2

10 to 195 mmHg

10 to 30 mmHg: 1 mmHg

 

 

 

1.0 to 26 kPa

1.0 to 4.0 kPa: 0.1 kPa

 

 

 

 

 

32 to 100 mmHg: 2 mmHg

 

 

 

 

 

4.2 to 13 kPa: 0.2 kPa

 

 

 

 

 

105 to 195 mmHg: 5 mmHg

 

 

 

 

 

13.5 to 26 kPa: 0.5 kPa

 

 

 

 

 

 

 

 

SvO2

SvO2 Performance Specifications

 

 

 

 

SvO2

Range

10% to 100%

 

 

 

Accuracy

± 2 % (i.e. ± 2 units), 1 standard deviation over 40 % to 100 %

 

 

 

 

range.

 

 

 

 

 

 

 

 

 

 

Resolution

1%

 

 

 

 

 

 

 

 

 

Stability (system)

 

 

Drift < 2% over 24 hours

 

 

 

 

 

 

 

Response Time (10 % to 90%)

5 seconds

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SvO2 Alarm Specifications

Range

 

Adjustment

Delay

SvO2

 

10% to 100%

 

1%

max. 15+4 seconds after value goes beyond the

 

 

 

 

 

low/high alarm limit settings

 

 

 

 

 

 

 

EEG

Complies with IEC 60601-2-26:1994/EN60601-2-26:1994.

363

Page 383
Image 383
Philips MP40/50, MP60/70/80/90, MP20/30 manual SvO2, SvO 2 Performance Specifications

MP60/70/80/90, MP40/50, MP20/30 specifications

Philips MP20/30, MP40/50, and MP60/70/80/90 are advanced patient monitoring systems designed to enhance clinical workflows and patient care in diverse healthcare settings. These devices combine cutting-edge technologies and user-friendly interfaces, making them essential tools for hospitals and healthcare providers.

The MP20 and MP30 are notable for their compact design, making them ideal for environments where space is limited. They come equipped with a variety of monitoring parameters, including ECG, SpO2, and non-invasive blood pressure, providing a comprehensive overview of patient health. These models are particularly suited for use in general wards or recovery areas, offering reliability and ease of use.

Moving to the MP40 and MP50, these monitors incorporate more advanced functionalities, including multi-gas monitoring, which is vital for anesthetic applications. They feature a robust data storage capacity, allowing for extensive patient history tracking and improved decision-making capabilities. The intuitive touchscreen interfaces facilitate quick access to essential information, enabling healthcare professionals to respond efficiently to changes in a patient's condition.

The MP60, MP70, MP80, and MP90 series represent the pinnacle of Philips' patient monitoring technology. These monitors are equipped with advanced algorithms that enhance the precision of vital sign measurements, including advanced ECG analysis capabilities that assist in detecting arrhythmias. The MP80 and MP90 models further integrate advanced monitoring technologies such as 3D visualization for hemodynamics and deep learning algorithms for early warning systems.

Connectivity is a key feature across the Philips monitoring range, with many models offering seamless integration with hospital information systems and electronic health records. This connectivity ensures that patient data is easily accessible to clinicians, promoting informed decision-making. Moreover, alarm management features are designed to minimize false alarms, significantly improving the staff's response time by focusing on critical alerts.

In summary, Philips MP20/30, MP40/50, MP60/70/80/90 monitors deliver exceptional monitoring capabilities, combining user-friendly design with advanced technology. Their varied features allow them to cater to different clinical needs, enhancing patient care and supporting healthcare professionals in their critical roles. With these monitors, healthcare facilities can achieve a new standard in patient monitoring, ensuring quality and safety in patient management.