Event Triggers

23 Event Surveillance

If you use alarm limits as event triggers, the event capture is triggered automatically when your patient’s values violate set alarm limits, or when a specified alarm condition, such as apnea, occurs. No events of this kind are triggered if alarms are switched off. Changing alarm limits changes the event trigger definitions.

The asterisk symbols beside the trigger tell you about the alarm triggers:

***indicates a high priority (red) alarm

**indicates a lower priority (yellow) alarm. This category includes short yellow arrhythmia alarms, which may be configured to be shown as one star alarms (*) on your monitor.

Triggers without asterisk symbols are user-defined triggers.

If you set user-defined threshold triggers, you can define event triggers that are independent of alarm limits. You must set a threshold value and a threshold time for the trigger. If you set the trigger threshold time to 12 seconds, the monitor triggers an event if the threshold is violated for more than 12 seconds.

If you set On Measurement triggers, you can set aperiodic measurements (such as NBP) or procedures (such as Cardiac Output) to trigger an event whenever they start.

If you set user-defined deviation triggers, you can define event triggers that are independent of specific limits and based instead on deviations from the current values. You must set a deviation and a period of time in which the deviation occurs. There are three types of deviation available: ANY deviation, UP deviation where only changes in a positive direction are detected and DOWN deviation where only changes in a negative direction are detected. The deviation can be defined either in relative terms as a percentage, for example 10%, or as an absolute value, such as 10 bpm.

Event Retriggering

If a condition that triggered an event persists and the values remain beyond the trigger threshold, a new event will not be triggered.

For a new event to be triggered by the same condition, the measured values for at least one of the triggers must cross back into the normal range and then recross the trigger threshold.

Event Notification

(Advanced Event Surveillance Only). You can be notified when an event is detected. For each event group you can define a type of notification depending on the severity of the event conditions. The notification can be a status message with a prompt tone or a standard *, **, or *** alarm notification. These event alarms are handled exactly like measurement alarms; they can be silenced and are also suspended when all alarms are suspended. You should only use alarm notification for events which are comparable in severity to standard measurement alarms to avoid potential confusion due to too many alarms. Notification in the form of an alarm is not available when the trigger condition is At Least One Param. Selecting None switches event notification off.

Setting the type of notification, or switching notification off, is done in Configuration Mode.

Setting Triggers for NER and Basic Event Surveillance

1Select Main Setup ->Event Surveill. ->Setup Events to enter the Event Setup window.

245

Page 265
Image 265
Philips MP20/30, MP40/50 Event Retriggering, Event Notification, Setting Triggers for NER and Basic Event Surveillance

MP60/70/80/90, MP40/50, MP20/30 specifications

Philips MP20/30, MP40/50, and MP60/70/80/90 are advanced patient monitoring systems designed to enhance clinical workflows and patient care in diverse healthcare settings. These devices combine cutting-edge technologies and user-friendly interfaces, making them essential tools for hospitals and healthcare providers.

The MP20 and MP30 are notable for their compact design, making them ideal for environments where space is limited. They come equipped with a variety of monitoring parameters, including ECG, SpO2, and non-invasive blood pressure, providing a comprehensive overview of patient health. These models are particularly suited for use in general wards or recovery areas, offering reliability and ease of use.

Moving to the MP40 and MP50, these monitors incorporate more advanced functionalities, including multi-gas monitoring, which is vital for anesthetic applications. They feature a robust data storage capacity, allowing for extensive patient history tracking and improved decision-making capabilities. The intuitive touchscreen interfaces facilitate quick access to essential information, enabling healthcare professionals to respond efficiently to changes in a patient's condition.

The MP60, MP70, MP80, and MP90 series represent the pinnacle of Philips' patient monitoring technology. These monitors are equipped with advanced algorithms that enhance the precision of vital sign measurements, including advanced ECG analysis capabilities that assist in detecting arrhythmias. The MP80 and MP90 models further integrate advanced monitoring technologies such as 3D visualization for hemodynamics and deep learning algorithms for early warning systems.

Connectivity is a key feature across the Philips monitoring range, with many models offering seamless integration with hospital information systems and electronic health records. This connectivity ensures that patient data is easily accessible to clinicians, promoting informed decision-making. Moreover, alarm management features are designed to minimize false alarms, significantly improving the staff's response time by focusing on critical alerts.

In summary, Philips MP20/30, MP40/50, MP60/70/80/90 monitors deliver exceptional monitoring capabilities, combining user-friendly design with advanced technology. Their varied features allow them to cater to different clinical needs, enhancing patient care and supporting healthcare professionals in their critical roles. With these monitors, healthcare facilities can achieve a new standard in patient monitoring, ensuring quality and safety in patient management.