Technical Alarm Messages (INOPs)

4 Patient Alarms and INOPs

INOP Message, Indication

Source

What to do

 

 

 

NBP INTERRUPTED

NBP

Check the tubing and cuff for leakages or kinks. Check that

Numeric is replaced by -?-

 

you are using the correct cuff size and placement, and that the

INOP tone.

 

correct patient category is selected. Try restarting the

 

 

measurement.

 

 

If the INOP occurs repeatedly, contact your service personnel.

 

 

You can silence this INOP, but the INOP message remains

 

 

visible until the next measurement is started or the Stop All

 

 

SmartKey is selected.

 

 

This INOP arises when the measurement needed longer than

 

 

the maximum time for inflation, deflation or the total

 

 

measurement.

NBP MEASURE FAILED

NBP

Check that you are using the correct cuff size and placement,

Numeric may be displayed with a -?-

 

and that the correct patient category is selected. Try restarting

INOP tone.

 

the measurement.

 

 

If numerics are displayed, the monitor is able to measure mean

 

 

only and the alarm source is set to S, D or S&D.

 

 

You can silence this INOP, but the INOP message remains

 

 

visible until the next measurement is started or the Stop All

 

 

SmartKey is selected.

 

 

Check the condition and suitability of the patient for NBP

 

 

monitoring. Use another cuff to continue measuring.

No Central Monit.

Monitor

There is a problem with the communication to the network.

INOP tone

 

Central monitoring is currently not possible (no patient alarms

 

 

or information). Check the connection. Contact your service

 

 

personnel.

NO PPV FROM MeasSrv

MMS or

The measurement server does not supply a beat-to-beat arterial

at Information Center

FMS

pressure value. Contact your service personnel.

NO PPV FROM <Server>

MMS or

The measurement server does not supply a beat-to-beat arterial

 

FMS

pressure value. Contact your service personnel.

P/P1/P2/P3/P4 INOPS

PRESS

See <Pressure label> INOPS (under Pressure).

PAP INOPS

PRESS

See <Pressure label> INOPS (under Pressure).

PPV BAD <Pressure Label>

PRESS

The arterial pressure source selected for PPV is not providing a

SIGNAL

 

pulsatile signal.

PPV BAD SIGNAL

PRESS

The arterial pressure source selected for PPV is not providing a

at Information Center

 

pulsatile signal.

PPV CHK SOURCES

PRESS

The arterial pressure source selected for PPV is unplugged or

 

 

switched off. When this INOP has displayed for 1 minute PPV

 

 

will be switched off.

<Pressure label> ARTIFACT

PRESS

A non-physiological event is detected (for example, a flush or

Numeric displayed with -?-

 

blood sample). A resulting high limit alarm will be suppressed.

<Pressure label>

PRESS

A Pressure measurement label in the measurement server or

DEACTIVATED

 

extension has been deactivated, either by connecting a Temp

INOP tone

 

transducer in the shared Press/Temp socket, or by deactivating

 

 

the label in the Measurement Selection window.

 

 

The measurement automatically disappears from the display.

 

 

To switch the measurement on again, either reconnect a

 

 

Pressure transducer or reactivate the measurement label in the

 

 

Measurement Selection window.

 

 

 

69

Page 89
Image 89
Philips MP40/50 NBP Interrupted, NBP Measure Failed, P1/P2/P3/P4 Inops, PPV BAD Signal, PPV CHK Sources, Deactivated

MP60/70/80/90, MP40/50, MP20/30 specifications

Philips MP20/30, MP40/50, and MP60/70/80/90 are advanced patient monitoring systems designed to enhance clinical workflows and patient care in diverse healthcare settings. These devices combine cutting-edge technologies and user-friendly interfaces, making them essential tools for hospitals and healthcare providers.

The MP20 and MP30 are notable for their compact design, making them ideal for environments where space is limited. They come equipped with a variety of monitoring parameters, including ECG, SpO2, and non-invasive blood pressure, providing a comprehensive overview of patient health. These models are particularly suited for use in general wards or recovery areas, offering reliability and ease of use.

Moving to the MP40 and MP50, these monitors incorporate more advanced functionalities, including multi-gas monitoring, which is vital for anesthetic applications. They feature a robust data storage capacity, allowing for extensive patient history tracking and improved decision-making capabilities. The intuitive touchscreen interfaces facilitate quick access to essential information, enabling healthcare professionals to respond efficiently to changes in a patient's condition.

The MP60, MP70, MP80, and MP90 series represent the pinnacle of Philips' patient monitoring technology. These monitors are equipped with advanced algorithms that enhance the precision of vital sign measurements, including advanced ECG analysis capabilities that assist in detecting arrhythmias. The MP80 and MP90 models further integrate advanced monitoring technologies such as 3D visualization for hemodynamics and deep learning algorithms for early warning systems.

Connectivity is a key feature across the Philips monitoring range, with many models offering seamless integration with hospital information systems and electronic health records. This connectivity ensures that patient data is easily accessible to clinicians, promoting informed decision-making. Moreover, alarm management features are designed to minimize false alarms, significantly improving the staff's response time by focusing on critical alerts.

In summary, Philips MP20/30, MP40/50, MP60/70/80/90 monitors deliver exceptional monitoring capabilities, combining user-friendly design with advanced technology. Their varied features allow them to cater to different clinical needs, enhancing patient care and supporting healthcare professionals in their critical roles. With these monitors, healthcare facilities can achieve a new standard in patient monitoring, ensuring quality and safety in patient management.