16 Monitoring tcGas

Setting the tcGas Sensor Temperature

Setting the tcGas Sensor Temperature

1In the Setup tcGas menu, select Transducer Temp.

2Choose a temperature value appropriate for your patient’s age, weight and physical condition in accordance with your hospital policy.

Usually, a higher transducer temperature gives a better correlation and a quicker response time. However, higher temperatures also increase the risk of skin burns. Most physicians prefer a temperature between 42°C (107° F) and 44°C (111° F), and a site time of four hours or less. Usually, the higher the transducer temperature, the less the site time should be. Whenever you change the temperature setting, the monitor forces you to make a new calibration.

Using the tcGas Site Timer

Availability and behavior of the site timer depend on your monitor’s configuration.

WARNING Prolonged continuous monitoring may increase the risk of undesirable changes in skin characteristics, such as irritation, reddening, blistering or burns. If the site timer is disabled, the transducer will heat indefinitely while on a patient. Change the site regularly, in accordance with medical procedures in your hospital.

Setting the tcGas Site Timer

The site timer helps reduce the risk of skin burn by ensuring that the transducer is used at one site for no longer than a predefined period. It reminds you when this period expires.

1In the Setup tcGas menu, select Site Time.

2Choose the time you want the transducer to remain on the measurement site. The optimum time depends on the transducer temperature and your patient’s skin sensitivity.

The site timer starts automatically when you remove the calibrated transducer from the calibration chamber. If you return the transducer to the chamber and then remove it again, the site time continues to count down the remaining time; it does not start a new time period. The time remaining before the site timer expires appears in the status line which is visible as long as the Setup tcGas menu is open. When the time expires, the monitor sounds a tone and a displays a change site INOP. The monitor either switches off the transducer heating or continues monitoring, depending on its configuration. Although you can reuse the transducer for up to two hours after the heating is switched off, without making a new calibration, you are recommended to recalibrate before applying it to a patient. After two hours without heat, you must recalibrate.

During the initial few minutes of use, the monitor eliminates false alarms by temporarily suppressing tcGas alarms. It displays the “STABILIZING” INOP. After you apply the transducer to the skin, the instrument reading slowly assumes a steady value. The reading stabilizes when the measurement site is warm and local hyperemization is complete. This takes 10 to 20 minutes for the tcpO2 reading and three to seven minutes for tcpCO2.

196

Page 216
Image 216
Philips MP60/70/80/90 manual Setting the tcGas Sensor Temperature, Using the tcGas Site Timer, Setting the tcGas Site Timer

MP60/70/80/90, MP40/50, MP20/30 specifications

Philips MP20/30, MP40/50, and MP60/70/80/90 are advanced patient monitoring systems designed to enhance clinical workflows and patient care in diverse healthcare settings. These devices combine cutting-edge technologies and user-friendly interfaces, making them essential tools for hospitals and healthcare providers.

The MP20 and MP30 are notable for their compact design, making them ideal for environments where space is limited. They come equipped with a variety of monitoring parameters, including ECG, SpO2, and non-invasive blood pressure, providing a comprehensive overview of patient health. These models are particularly suited for use in general wards or recovery areas, offering reliability and ease of use.

Moving to the MP40 and MP50, these monitors incorporate more advanced functionalities, including multi-gas monitoring, which is vital for anesthetic applications. They feature a robust data storage capacity, allowing for extensive patient history tracking and improved decision-making capabilities. The intuitive touchscreen interfaces facilitate quick access to essential information, enabling healthcare professionals to respond efficiently to changes in a patient's condition.

The MP60, MP70, MP80, and MP90 series represent the pinnacle of Philips' patient monitoring technology. These monitors are equipped with advanced algorithms that enhance the precision of vital sign measurements, including advanced ECG analysis capabilities that assist in detecting arrhythmias. The MP80 and MP90 models further integrate advanced monitoring technologies such as 3D visualization for hemodynamics and deep learning algorithms for early warning systems.

Connectivity is a key feature across the Philips monitoring range, with many models offering seamless integration with hospital information systems and electronic health records. This connectivity ensures that patient data is easily accessible to clinicians, promoting informed decision-making. Moreover, alarm management features are designed to minimize false alarms, significantly improving the staff's response time by focusing on critical alerts.

In summary, Philips MP20/30, MP40/50, MP60/70/80/90 monitors deliver exceptional monitoring capabilities, combining user-friendly design with advanced technology. Their varied features allow them to cater to different clinical needs, enhancing patient care and supporting healthcare professionals in their critical roles. With these monitors, healthcare facilities can achieve a new standard in patient monitoring, ensuring quality and safety in patient management.