34 Installation and Specifications Measurement Specifications
362
tcGas

Complies with IEC 60601-2-23:1999/EN60601-2-23:2000.

ITBV Range 50 to 6000 ml
Accuracy + 10% or 30 ml
Repeatability + 5% or 20 ml
CCO Range 0.1 to 25.0 l/min
Standard Deviation 10% or 0.3 l/min
Display Update 2 seconds nominal
Blood Temperature Range 17,0 to 43°C (62,6 to 109,4°F)
Injectate Temperature Range -1 to 30°C
C.O./CCO Alarm Specifications Range Adjustment Delay
TBlood 17 to 43°C Steps of 0.5°C (17 to 35°C)
Steps of 0.1°C (35 to 43°C)
Steps of 1°F (63 to 95°C)
Steps of 0.2°F (95 to 109°C
10 seconds after the value
exceeds the set limit range
CCO 0.1 to 25.0 l/min 0.1 l/min (0.1 to10.0 l/min)
0.5 l/min (10.0 to 25.0 l/min)
10 seconds after the value
exceeds the set limit range

C.O./CCO Performance Specifications

tcGas Performance Specifications

tcpO2Range 0 to 750 mmHg (0 to 100 kPa)
Accuracy 0.5% (+ 1 digit)
Resolution 1 mmHg (0.1 kPa)
Overall Accuracy
(incl. transducer)
0 to 160 mmHg + 5 mmHg
0.0 to 21.3 kPa + 0.7 kPa
>160 mmHg (21.3 kPA) + 3% of reading
tcpO2 Temperature Drift < 0.1% / °C
tcpO2 Te s t Si gn al 60 mmHg (8.0 kPa)
tcpO2 Drift, including transducer < 1 mmHg / h @ 10% O2
tcpO2 response time, including transducer < 30 s
tcpCO2Range 5 to 200 mmHg (0.7 to 26.7 kPa)
Accuracy 1.0% (1 + digit)
Resolution 1 mmHg (0.1 kPa)
Overall Accuracy
(incl. transducer)
0 to 76 mmHg + 5 mmHg
0.0 to 10.1 kPa + 0.7 kPa
>76 mmHg (10.1 kPa) + 10% of reading
tcpCO2 Temperature Drift < 0.1% / °C
tcpCO2 Test Signal 40 mmHg (5.3 kPa)
tcpCO2 Drift, including transducer < 1 mmHg / h @ 5% CO2