Technical Alarm Messages (INOPs)

4 Patient Alarms and INOPs

INOP Message, Indication

Source

What to do

 

 

 

CO2 CHECK CAL

CO2

The CO2 value is outside the measurement range. Perform an

Numeric is replaced by a - ? -

 

accuracy check for both calstick cells and, if necessary,

INOP tone.

 

recalibrate the transducer.

C.O. EQUIP MALF

C.O.

There is a problem with the C.O. hardware. Contact your

Numeric is replaced by a - ? -

 

service personnel.

INOP tone.

 

 

CO2 EQUIP MALF

CO2

The Measurement Server Extension is faulty. Unplug and

Numeric is replaced by - ? -

 

replug the Measurement Server with Extension. If you are using

INOP tone.

 

the mainstream method, unplug and replug the transducer or

 

 

try another transducer. If the INOP persists, contact your

 

 

service personnel.

CO2 FAILED CAL

CO2

Make sure that the Cal cell was changed between CAL1 and

Numeric is replaced by -?-

 

CAL2. Repeat the calibration. If the INOP reappears, try

INOP tone.

 

another transducer. If the INOP persists, contact your service

 

 

personnel.

CO2 NO SENSOR

CO2

There is no CO2 sensor connected. If you silence this INOP

Numeric is replaced by - ? -

 

the CO2 measurement will be switched off.

INOP tone.

 

 

CO2 NO TRANSDUC

CO2

There is no CO2 transducer connected. If you replace the

Numeric is replaced by - ? -

 

transducer, the new transducer must be calibrated. If you

INOP tone.

 

silence this INOP the CO2 measurement will be switched off.

CO2 NO TUBING

CO2

Either the FilterLine is disconnected, or an incorrect line is

Numeric is replaced by - ? -

 

attached. Check the connection. If necessary, connect another

INOP tone.

 

MicrostreamFilterline (only Microstream accessories can be

 

 

used).

 

 

If you silence this INOP, the measurement will be switched off.

CO2 OCCLUSION

CO2

The FilterLine or exhaust tube is blocked. Check the FilterLine

Numeric is replaced by a - ? -

 

and exhaust tube, then disconnect and reconnect the

INOP tone.

 

FilterLine. If the INOP persists, connect a new FilterLine.

CO2 OVERRANGE

CO2

The CO2 value is higher than the measurement range. If you

Numeric is replaced by - ? -

 

suspect a false high value, contact your service personnel.

INOP tone.

 

 

CO2 PUMP OFF

CO2

The pump has been switched off for fifteen minutes. To switch

Numeric is replaced by a - ? -.

 

it on again, select Pump On in the Setup CO2 menu.

CO2 PURGING

CO2

The Filterline is being purged to remove an occlusion in the

Numeric is replaced by a - ? -

 

line or airway adapter. If the occlusion is removed, the INOP

INOP tone.

 

will disappear. If not, the INOP CO2 OCCLUSION is

 

 

displayed.

CO2 SENSOR WARMUP

CO2

Wait until the sensor reaches operating temperature and the

Numeric is displayed with a - ? -

 

INOP disappears.

Microstream CO2: INOP tone.

 

 

Mainstream CO2: no INOP tone

 

 

C.O. UNPLUGGED

C.O.

Plug in the C.O. module. Silencing this INOP switches off the

numeric is replaced by -?-

 

measurement.

INOP tone.

 

 

CO2 UPDATE FW

CO2

The software in the Measurement Server Extension does not

Numeric is replaced by a - ? -

 

match the software in the MMS. Contact your service

INOP tone.

 

personnel.

CO2 WAIT CAL2

CO2

Calibration on the first calstick cell is complete. Place the

Numeric is replaced by a - ? -

 

transducer on the other calstick cell and start the CAL2

 

 

calibration cycle.

 

 

 

65

Page 85
Image 85
Philips MP20/30, MP40/50 CO2 Check CAL, CO2 Equip Malf, CO2 Failed CAL, CO2 no Sensor, CO2 no Transduc, CO2 no Tubing

MP60/70/80/90, MP40/50, MP20/30 specifications

Philips MP20/30, MP40/50, and MP60/70/80/90 are advanced patient monitoring systems designed to enhance clinical workflows and patient care in diverse healthcare settings. These devices combine cutting-edge technologies and user-friendly interfaces, making them essential tools for hospitals and healthcare providers.

The MP20 and MP30 are notable for their compact design, making them ideal for environments where space is limited. They come equipped with a variety of monitoring parameters, including ECG, SpO2, and non-invasive blood pressure, providing a comprehensive overview of patient health. These models are particularly suited for use in general wards or recovery areas, offering reliability and ease of use.

Moving to the MP40 and MP50, these monitors incorporate more advanced functionalities, including multi-gas monitoring, which is vital for anesthetic applications. They feature a robust data storage capacity, allowing for extensive patient history tracking and improved decision-making capabilities. The intuitive touchscreen interfaces facilitate quick access to essential information, enabling healthcare professionals to respond efficiently to changes in a patient's condition.

The MP60, MP70, MP80, and MP90 series represent the pinnacle of Philips' patient monitoring technology. These monitors are equipped with advanced algorithms that enhance the precision of vital sign measurements, including advanced ECG analysis capabilities that assist in detecting arrhythmias. The MP80 and MP90 models further integrate advanced monitoring technologies such as 3D visualization for hemodynamics and deep learning algorithms for early warning systems.

Connectivity is a key feature across the Philips monitoring range, with many models offering seamless integration with hospital information systems and electronic health records. This connectivity ensures that patient data is easily accessible to clinicians, promoting informed decision-making. Moreover, alarm management features are designed to minimize false alarms, significantly improving the staff's response time by focusing on critical alerts.

In summary, Philips MP20/30, MP40/50, MP60/70/80/90 monitors deliver exceptional monitoring capabilities, combining user-friendly design with advanced technology. Their varied features allow them to cater to different clinical needs, enhancing patient care and supporting healthcare professionals in their critical roles. With these monitors, healthcare facilities can achieve a new standard in patient monitoring, ensuring quality and safety in patient management.