Technical Alarm Messages (INOPs) 4 Patient Alarms and INOPs
65
CO2 CHECK CAL
Numeric is replaced by a - ? -
INOP tone.
CO2The CO2 value is outside the measurement range. Perform an
accuracy check for both calstick cells and, if necessary,
recalibrate the transducer.
C.O. EQUIP MALF
Numeric is replaced by a - ? -
INOP tone.
C.O. There is a problem with the C.O. hardware. Contact your
service personnel.
CO2 EQUIP MALF
Numeric is replaced by - ? -
INOP tone.
CO2The Measurement Server Extension is faulty. Unplug and
replug the Measurement Server with Extension. If you are using
the mainstream method, unplug and replug the transducer or
try another transducer. If the INOP persists, contact your
service personnel.
CO2 FAILED CAL
Numeric is replaced by -?-
INOP tone.
CO2Make sure that the Cal cell was changed between CAL1 and
CAL2. Repeat the calibration. If the INOP reappears, try
another transducer. If the INOP persists, contact your service
personnel.
CO2 NO SENSOR
Numeric is replaced by - ? -
INOP tone.
CO2There is no CO2 sensor connected. If you silence this INOP
the CO2 measurement will be switched off.
CO2 NO TRANSDUC
Numeric is replaced by - ? -
INOP tone.
CO2There is no CO2 transducer connected. If you replace the
transducer, the new transducer must be calibrated. If you
silence this INOP the CO2 measurement will be switched off.
CO2 NO TUBING
Numeric is replaced by - ? -
INOP tone.
CO2Either the FilterLine is disconnected, or an incorrect line is
attached. Check the connection. If necessary, connect another
Microstream Filterline (only Microstream accessories can be
used).
If you silence this INOP, the measurement will be switched off.
CO2 OCCLUSION
Numeric is replaced by a - ? -
INOP tone.
CO2The FilterLine or exhaust tube is blocked. Check the FilterLine
and exhaust tube, then disconnect and reconnect the
FilterLine. If the INOP persists, connect a new FilterLine.
CO2 OVERRANGE
Numeric is replaced by - ? -
INOP tone.
CO2The CO2 value is higher than the measurement range. If you
suspect a false high value, contact your service personnel.
CO2 PUMP OFF
Numeric is replaced by a - ? -.
CO2The pump has been switched off for fifteen minutes. To switch
it on again, select Pump On in the Setup CO2 menu.
CO2 PURGING
Numeric is replaced by a - ? -
INOP tone.
CO2The Filterline is being purged to remove an occlusion in the
line or airway adapter. If the occlusion is removed, the INOP
will disappear. If not, the INOP CO 2 OCCLUS ION is
displayed.
CO2 SENSOR WARMUP
Numeric is displayed with a - ? -
Microstream CO2: INOP tone.
Mainstream CO2: no INOP tone
CO2Wait until the sensor reaches operating temperature and the
INOP disappears.
C.O. UNPLUGGED
numeric is replaced by -?-
INOP tone.
C.O. Plug in the C.O. module. Silencing this INOP switches off the
measurement.
CO2 UPDATE FW
Numeric is replaced by a - ? -
INOP tone.
CO2The software in the Measurement Server Extension does not
match the software in the MMS. Contact your service
personnel.
CO2 WAIT CAL2
Numeric is replaced by a - ? -
CO2Calibration on the first calstick cell is complete. Place the
transducer on the other calstick cell and start the CAL2
calibration cycle.