Section 5

Checks And Adjustments

This section consists of two separate procedures. The first, a Performance Check, is used to determine compliance with the Performance Requirements in the Specification. The second is the Adjustment Procedure, that provides the instructions on how to adjust the instrument and return it to operation within the specification.

In both procedures, front- and rear-panel controls and connectors, on the instrument under test, are fully capitalized (e.g., 300 kHz RESOLUTION). Control and connector names on test equipment and internal controls and adjustments for the instrument under test are initial capitalized (e.g., Time/Div, Geometry, etc.).

Limits, tolerances, and waveforms given in this section are guides to adjustments and checks. They are not instrument specifications, except when listed in the Performance Requirement column of the Specification Tables in Section 1 of this manual.

Recommended Equipment List

The following equipment and accessory items are required to do the Performance Check and/or Adjustment Procedures. Broad specifications are followed by a piece of equipment that meets these specifications; in most cases, the recom- mended instrument was used in preparing the procedures that follow.

Electrical Instruments

1.Test Oscilloscope Vertical Amplifier:

30 MHz Bandwidth, 1 mV Sensitivity. Time Base:

10 ns/Division to 5 ms/Division sweep speeds, triggering to 5 MHz.

For example: a TEKTRONIX 7603 Oscilloscope with a 7A18 Dual-Trace Amplifier and a 7B53A Dual Time Base, or a TEKTRONIX 11403A Oscilloscope with a 11A34V Video Amplifier and an 11T5H Video Trigger. Also a 10X probe, P6106 (Tektronix Part No. 010-6106-03).

1705A Spectrum Monitor

5- 1

Page 96
Image 96