Checks and Adjustments

2.Leveled Sine Wave Generator, at least 250 kHz to 95 MHz.

For example: A TEKTRONIX SG503 Leveled Sine Wave Generator installed in a TEKTRONIX TM500 Series Power Module.

3.Voltmeter

Range, 0 to greater than 100 Vdc; accuracy, ±0.1%.

For example: A TEKTRONIX DM501A in a TM500 Series Power Module.

4.Power Module for powering and housing TEKTRONIX DM501A, DC503A, FG503, 067-0916-00, and 015-0408-00.

For example: A TEKTRONIX TM506 Power Module.

5.UHF Signal Generator

A frequency range of at least 900 to 1800 MHz, with an amplitude of -20 dBm, or more, and flatness within 3 dB over the frequency range. Accurate step attenuator calibrated in dB (0 to -60).

For example: A TEKTRONIX TR502 Tracking Generator and associated equipment, a Wavetek Model 3520, or a Hewlett Packard 8614A Signal Generator with 8496A Attenuator/110 dB.

6.Variable Autotransformer

For example: General Radio Metered Auto Transformer W10MT3W. If

220 volt operation must be checked, a conversion transformer or appropriate 220 volt autotransformer is needed.

7.Comb Generator

For example: Tektronix Part No. 015-1054-00.

Auxiliary Equipment

8.50Ω Coaxial Cable Two required.

For example: Tektronix Part No. 012-0057-01.

9.Bnc Male-to-bnc Male Adapter

For example: Tektronix Part No. 103-0029-00.

10.F-Type Male-to-bnc Female Adapter

For example: Tektronix Part No. 103-0158-00.

5- 2

1705A Spectrum Monitor

Page 97
Image 97
Tektronix 1705A instruction manual Auxiliary Equipment