Maintenance

Table 6- 7: Test Selectable Components

Circuit Number

Nominal Value

Range of Values

Selection Criteria

 

 

 

 

R50

5.62K

2.7K to 8.2K

To correct nonlin-

R74

3.40K

1K to 6.8K

earities of L--band

R73

18.2K

18K to 27K

tuners.

 

 

 

 

R53

115.0K

115K to 130K

To alter magnifier

 

 

 

range when ad-

 

 

 

justment range of

 

 

 

A3R38 is insuffi-

 

 

 

cient.

 

 

 

 

Mechanical Disassembly/Assembly

Use these instructions for disassembly and reverse them for reassembly, unless otherwise noted.

WARNING. Before attempting any disassembly/assembly of the instrument, be sure to disconnect the power cord.

CAUTION. Do not reinsert screws in the rear panel when the instrument is removed from the cabinet.

NOTE. All screws, unless otherwise noted, are TORX￿ screws and can be removed with a T15 screwdriver tip (Tektronix part number 003-0966-00). The exception is #2 Pozidrive￿ screws which can be removed with a #1 Pozidrive￿ tip (003-0443-00).

Do not reinsert screws in the rear panel when the instrument is removed from the cabinet.

Bezel Removal

1.Remove the two bezel screws. See Figure 6-8.

2.Grasping the bottom of the bezel, pull out and upward. There are two hinges at the top of the bezel that hold it in place; once the bezel is at an approxi- mate 45° angle with the front panel, they will disengage.

1705A Spectrum Monitor

6- 19

Page 144
Image 144
Tektronix 1705A instruction manual Mechanical Disassembly/Assembly, Bezel Removal, Maintenance Test Selectable Components