Checks and Adjustments
5-4 1705A Spectrum Monitor
6. Check Resolution Filter
b. CHECK --that the marker width 0.6 Divisions down from the top is
3 Divisions, ±1 Division.
e. CHECK --that the marker width 0.6 Divisions down from the top of the
signal is 2 minor Divisions.
f. CHECK --that the 10 kHz marker amplitude matches the highest point
of the 300 kHz marker amplitudewithin 1 minor Division.
7. Check 70 MHz Gain and Flatness
e. CHECK --for a marker amplitude change of 2 Divisions, ±2 minor
Divisions when switching the leveled sine wave generatorAmplitude
Multiplier between.01 and 0.1 (a 20 dB change).
i. CHECK --that the tip of the marker is on the crt center line, ±1 minor
Division (vertically).
8. Check VideoFilter
d. CHECK --that the baseline noise amplitude drops approximately 50%
when the VIDEO FILTERis ON.
9. Check Sweep Speed
b. CHECK --for a solid trace with almost no flicker.
d. CHECK --for approximately 3 to 4 sweeps per second.
Alternate Method
e. CHECK --for a ramp duration of 20 ms ±10 ms.
h. CHECK --for a ramp duration of 200 ms ±100 ms.
10. Check L--BandLin earity
e. CHECK --for 10 frequency markers.
f. CHECK --that each marker is within 1 min or Division of a major
graticuleline.
11. Check L--BandSPAN/DIV and Readout
d. CHECK --for 1 marker every 5 Divisions, ±1 Division.
i. CHECK --for 1 marker every 5 Divisions, ±1 Division.
o. CHECK --that the marker is int ensified.
q. CHECK --that the marker is on s creen.