Maintenance
6-16 1705A Spectrum Monitor
Table6- 5: High Volts Supply Fault Symptoms
Symptom Procedure
Unable to focus crt using the front-
panel control
Focus Amplifier Check
Unable to adjust crt intensity using the
front-panel control
Z--Axis Amplifier Check
Grid Drive Check
No crt display High Voltage Oscillator Check
CRT Voltage Check
Load the Low VoltsSupply with the instrument, or with the 20resistor as
stated at thebeginning of the Troubleshooting Procedure.
Focus AmplifierC heck
Unsolder and lift one end of R24.
Power up the Power Supply.
Using the digital multimeter,measure the voltage between TP1 and the collector
of Q1. It should be approximately--140 V.
Reconnect thelifted end of R24.
Z-Axis Amplifier Check
Unsolder and lift one end of R8.
Power up the Power Supply.
Using the digital multimeter,measure the voltage between TP1 and the collector
of Q4. It should be approximately+10 V.
Short together the base and emitterof Q5. The collector of Q4 should be
approximately+100 V.
Reconnect thelifted end of R8.
Grid Drive Check
Turnoff the Power Supply. Use the digital multimeters diode check to test
CR1, CR2, CR3, CR5,and CR6 for shorts.
Power up the Power Supply.
Using the digital multimeter,measure the voltage between TP1 and the cathode
of CR5. It should vary betweenapproximately +75 and +200 V as R58 (CRT
Bias) is adjusted.