Syntax and Commands
3-34 AFG3000 Series Arbitrary/Function Generators Reference Manual
OUTPut:TRIGger:MODE
This command sets or queries the mode (trigger or sync) for Trigger Output signal.
When the burst count is set to Inf-Cycles in burst mode, TRIGger indicates that the
infinite number of cycles of waveform will be output from the Trigger Output
connector.
When the burst count is set to Inf-Cycles in burst mode, SYNC indicates that one
pulse waveform is output from the Trigger Output connector when the Inf-Cycles
starts.
When Run Mode is specified other than Burst Inf-Cycles, TRIGger and SYNC
have the same effect.
Group Output
Syntax OUTPut:TRIGger:MODE {TRIGger|SYNC}
OUTPut:TRIGger:MODE?
Arguments TRIGger means TRIGger is selected for Trigger Out.
SYNC means SYNC is selected for Trigger Out.
Returns TRIG|SYNC
Examples OUTPut:TRIGger:MODE SYNC
outputs one cycle waveform from the Trigger Output connector when Inf-Cycles
starts.