Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual 3-81
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe]This command enables or disables the function to copy the voltage level of one
channel to another channel.
The[SOURce[1|2]]:VOLTage:CONCurrent[:STATe] command copies the
voltage level of the channel specified by the header suffix to another channel. If you
specify CH 1 with the header, the CH 1 voltage level will be copied to CH 2.
The query command returns “0” (off) or “1” (on).
If your arbitrary/function generator is a single-channel model, this command is not
supported.
Group Source
Syntax [SOURce[1|2]]:VOLTage:CONCurrent[:STATe] {ON|OFF|<NR1>}
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe]?
Arguments ON or <NR1>≠0 enables the concurrent copy function.
OFF or <NR1>=0 disables the concurrent copy function.
Returns <NR1>
Examples SOURce1:VOLTage:CONCurrent:STATe ON
enables the concurrent copy function.