Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual 3-77
[SOURce]:ROSCillator:SOURceThis command sets the reference clock to either internal or external.
Group Source
Syntax [SOURce]:ROSCillator:SOURce {INTernal|EXTernal}
[SOURce]:ROSCillator:SOURce?
Arguments INTernal means that the reference clock is set to internal.
EXTernal means that the reference clock is set to external.
Returns INT|EXT
Examples SOURce:ROSCillator:SOURce INTernal
selects the internal clock reference.
[SOURce[1|2]]:SWEep:HTIMeThis command sets or queries the sweep hold time. Hold time represents the
amount of time that the frequency must remain stable after reaching the stop
frequency.
Group Source
Syntax [SOURce[1|2]]:SWEep:HTIMe {<seconds>|MINimum|MAXimum}
[SOURce[1|2]]:SWEep:HTIMe? {MINimum|MAXimum}
Arguments <seconds>::=<NRf>[<units>]
where
<NRf> is the hold time in seconds.
<units>::=[ns | ยตs | ms | s]
Returns <seconds>
Examples SOURce1:SWEep:HTIMe 1ms
sets the CH 1 hold time to 1 ms.