Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual 3-79
[SOURce[1|2]]:SWEep:RTIMeThis command sets or queries the sweep return time. Return time represents the
amount of time from stop frequency through start frequency. Return time does not
include hold time.
Group Source
Syntax [SOURce[1|2]]:SWEep:RTIMe {<seconds>|MINimum|MAXimum}
[SOURce[1|2]]:SWEep:RTIMe? {MINimum|MAXimum}
Arguments <seconds>::=<NRf>[<units>]
where
<NRf> is the return time in seconds.
<units>::=[ns | ยตs | ms | s]
Returns <seconds>
Examples SOURce1:SWEep:RTIMe 1ms
sets the CH 1 return time to 1 ms.
[SOURce[1|2]]:SWEep:SPACingThe [SOURce[1|2]]:SWEep:SPACing command selects linear or logarithmic
spacing for the sweep for the specified channel.
The query command returns the type for the sweep spacing for the specified
channel.
Group Source
Syntax [SOURce[1|2]]:SWEep:SPACing {LINear|LOGarithmic}
[SOURce[1|2]]:SWEep:SPACing?
Arguments LINear sets the sweep spacing to linear.
LOGarithmic sets the sweep spacing to logarithmic.
Returns LIN|LOG
Examples SOURce1:SWEep:SPACing LINear
sets the CH1 sweep spacing to linear.