Syntax and Commands
3-86 AFG3000 Series Arbitrary/Function Generators Reference Manual
[SOURce[1|2]]:VOLTage:LIMit:HIGH
This command sets or queries the higher limit of the output amplitude high level
for the specified channel. If your instrument is a dual-channel model and the
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe] command is set to ON, then the
higher level limit of the other channel is the same value.
Group Source
Related Commands [SOURce[1|2]]:VOLTage:CONCurrent[:STATe]
Syntax [SOURce[1|2]]:VOLTage:LIMit:HIGH {<voltage>|MINimum|MAXimum}
[SOURce[1|2]]:VOLTage:LIMit:HIGH? {MINimum|MAXimum}
Arguments <voltage>::=<NRf>[<units>]
where
<NRf> is the higher limit of output amplitude.
<units>::=[mV | V]
Returns <voltage>
Examples SOURce1:VOLTage:LIMit:HIGH 1V
sets the higher limit of CH 1 output amplitude to 1 V.