Syntax and Commands
3-92 AFG3000 Series Arbitrary/Function Generators Reference Manual
STATus:QUEStionable:CONDition?This query-only command returns the contents of the Questionable Condition
Register.
Group Status
Syntax STATus:QUEStionable:CONDition?
Arguments None
Returns <bit_value>::=<NR1>
Examples STATus:QUEStionable:CONDition?
might return 32 which indicates that the QCR contains the binary number
00000000 00100000 and the accuracy of frequency is questionable.
STATus:QUEStionable:ENABleThis command sets or queries the mask for the Questionable Enable Register.
Group Status
Syntax STATus:QUEStionable:ENABle <bit_value>
STATus:QUEStionable:ENABle?
Arguments <bit_value>::=<NR1>
Returns <bit_value>
Examples STATus:QUEStionable:ENABle 32
sets the FREQuency bit in the QENR to on.