Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual 3-67
[SOURce[1|2]]:PULSe:DCYCleThis command sets or queries the duty cycle of the pulse waveform for the speci-
fied channel. The setting range is 0.1% to 99.9% in increments of 0.1.
The arbitrary/function generator will hold the settings of leading edge and trailing
edge when the duty cycle is varied.
Refer to the [SOURce[1|2]]:PULSe:WIDTh command for the setting range.
Group Source
Related Commands [SOURce[1|2]]:PULSe:WIDTh
Syntax [SOURce[1|2]]:PULSe:DCYCle {<percent>|MINimum|MAXimum}
[SOURce[1|2]]:PULSe:DCYCle? {MINimum|MAXimum}
Arguments <percent>::=<NR2>[<units>]
where
<NRf> is the duty cycle.
<units>::=PCT
Returns <percent>
Examples SOURce1:PULSe:DCYCle 80.5
sets the duty cycle of the pulse waveform on CH 1 to 80.5%.