Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual 3-17
CALibration[:ALL]
The CALibration[:ALL] command performs an internal calibration.
The CALibration[:ALL]? command performs an internal calibration and returns
0 (Pass) or a calibration error code.
Group Calibration and Diagnostic
Related Commands *CAL?
Syntax CALibration[:ALL]
CALibration[:ALL]?
Arguments None
Returns <NR1>
Examples CALibration:ALL
performs an internal calibration.
CALibration:ALL?
performs an internal calibration and returns results. For example, it might return 0,
which indicates that the calibration completed without any errors.
*CLS (No Query Form)
This command clears all the event registers and queues, which are used in the arbi-
trary/function generator status and event reporting system.
Group Status
Syntax *CLS
Arguments None
Examples *CLS
clears all the event registers and queues.
NOTE. The self-calibration can take several minutes to complete. During this time,
the arbitrary/function generator does not execute any commands. Do not power off
the instrument during the self-calibration.